What does one eV equal to
|
1.6 x 10 e -19 Joules
|
|
What is the charge of an electron?
|
1. 6 e -19 C
|
|
What is the unit of electric current?
|
Ampere = charge / unit of time = coulomb / sec
|
|
What is power?
|
Rate of performing work
|
|
What is the unit of power?
|
Watt = joules / sec = volts x ampere = current x voltage
|
|
What's the exposure time for a typical chest x-ray?
|
10 mS
|
|
What's the unit of frequency and what does it mean?
|
Hertz = cycles per second
|
|
Velocity of photons is equal to ------ x ------
|
wavelength x frequency
|
|
What is the usual target in the x-ray tube?
|
Tungsten
|
|
What is the usual target in the mammography unit?
|
Molybdinium or Rhodium
|
|
What is Brehmsstralung radiation?
|
Fraction of the electron kinetic energy is emitted as an x-ray photon, and the electron travels through the target atom, and comes out as an electron of lesser kinetic energy.
|
|
How does Brehnmsstralung radiation vary with increasing target atomic number?
How does Brehmsstralung radiation vary with increasing Voltage |
Brehmsstralung radiation increases with increasing target atomic number and accelerating voltage
|
|
When is characteristic radiation produced?
|
When incoming electrons eject atomic electrons, usually occurs as k shell electrons are ejected, incident electron energy has to be greater than the k shell electron energy.
|
|
What three particles are ejected from characteristic radiation?
|
K shell electron
Possibly an electron with lower energy - Auger electron x- ray photon |
|
What is the characteristic photon energy from tungsten?
|
58 - 67 keV
|
|
What is the characteristic energy from Molybdinium?
|
17 to 19 keV
|
|
Approximately what percent do the K shell characteristic energy contribute to the x-ray spectrum at 100 KeV?
|
10%
|
|
How is x-ray beam intensity related to the exposure time and tube current?
|
Directly to ms of exsposure and mA of tube current
|
|
What is the x-ray beam intensity measured in?
|
Air kerma
|
|
Does increasing mAs change the energy of x-ray spectrum?
|
No, it does not change the energy spectrum
|
|
How is the x-ray beam intensity related to voltage output?
|
intensity : kV (squared)
|
|
In radiographic imaging, if KV is increased by 10 - 15% or increased by 10 kV, what effect does it have on air kerma (x-ray intensity) equivalent to?
|
Doubling the mAs
|
|
What are x-ray windows in mammography made of and why?
|
Berryllium, which absorbs less of lower energy x-rays
|
|
What is line focus principle
|
Permits larger heat loading while minimizing the size of the focal spot (due to anode angle, the thickness of the focal spot of the x-rays is less than the thickness of the electrons).
|
|
What is the anode angle and why is it important?
What are the typical anode angle ranges? |
Anode angle is the angle between incident electrons and the anode target. It determines the focal spot size.
7 - 20 degrees |
|
Why do focal spots need to be small?
Why do focal spots need to be large? |
Small - to produce sharper images
Large - to permit heat loading, when short exposure time is important |
|
What is the nominal focal spot for:
Magnification mammography Mammography Flouroscopy Radiographs |
Mag mammo: 0.1 mm
Mammo: 0.3 mm Fluoroscopy: 0.6 mm Radiographs: 1.2 mm |
|
What does an operator choose in automatic exposure mode and what is controlled by the generator circuit?
|
Operator chooses the KeV whereas generator circuit controls the tube current and exposure time
|
|
What is the typical tube current for?
Radiographs fluoroscopy |
radiographs: 100 - 1000 mA
Fluorscopy: 1 - 5 mA |
|
What is the typical voltage for radiographs and fluoroscopy?
|
70 - 90 KV
|
|
What is the typical voltage for abdominal radiograph?
What is the typical voltage for chest radiograph? CT? |
80 kV
140 kV 120 kV |
|
What does energy in Joules equal which is deposited on an anode?
|
V x I x t Joules = W x t Joules = Power x time = voltage x current x time
|
|
What is tube rating?
What's typical tube rating of x-ray tube with small focal spot? What's typical tube rating of x-ray tube with large focal spot? Tube rating of fluoroscopy? |
Maximum allowable power (kW) at an exposure time of 0.1 sec.
Small focal spot x-ray tube rating: 25 kW Large focal spot x-ray tube rating: 100 kW fluoroscopy tube rating 100 - 500 kW |
|
What is the maximum allowable amount for leakage radiation?
What is leakage radiation |
1 mGy / hour at a distance of 1 m from the x-ray tube?
Leakage radiation is the radiation that escapes the x-ray tube housing. |
|
What's the typical anode angle for radiography?
|
7 - 20 degrees
|
|
What is the binding energy of K shell electrons in tungsten?
|
70 kEV , an electron with energy of 65 keV won't be able to knock out any electrons at all.
|
|
What is the typical x-ray tube filament current?
|
4 A or 4000 mA (careful with the units)
|