• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/47

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

47 Cards in this Set

  • Front
  • Back
  • 3rd side (hint)
What condition must be met in order to maintain virtually zero resistance in a superconducting magnet?
The superconducting windings must be maintained at a temperature approaching absolute zero. (or 4 degree K)

*This can be achieved by bathing it in liquid helium.
At this temperature the wire achieves near zero electrical resistance which allows huge currents to flow through it thereby generating the magnetic field.
If a voltage source was connected to a solenoid, so that the current flowed as in the diagram below, how would a magnetic field be generated?
To the right.
Right hand rule
*Rule states that within the fingers of the right hand curled so that the finger tips point in the direction of the flow of current through a coil of wire then the extended thumb points in the direction of the resulting magnetic field.
The purpose of the shims is to:
Make the magnetic field more uniform.

*Provides homogeneity of the main magnetic field thereby removing most of the field inhomogeneities.
The purpose of magnetic shielding is to:
Keep the fringe magnetic field closer to the magnet.

*Note: Magnetic shielding is not designed to control RF energy and does not act as a Faraday shield.
What keeps the RF energy used in scanning from interfering with the external environment and keeps the stray radio signals in the environment from creating artifacts in the MR images?
RF or Faraday shielding
The correct location of the z-gradient coils in a superconductive MRI system is wrapped ______________________ a cylinder within the bore of the magnet.
Completely around each end of

*The X gradient is generated by coils wrapped on the top and bottom of the cylinder.
***There is no gradient coil which is wrapped in a spiral fashion. Although tha main magnetic field can be generated in this way.
Which of the following is affected by the transmitter bandwidth?
The slice thickness
Which of the following MRI system components is responsible for the timing and correct performance of each component once the sequence has been started?
Pulse sequence controller
What performs calculations on MR echo and image data?
The array processor
Which of the following describes the operational mechanism of an electromagnet?
A power supply is connected to the ends of a coil of long wire, generating a strong magnetic field.
Gradient rise time is defined as which of the following?
The time required for a gradient to reach its specified amplitude beginning from a resting state.
Which of the following describes the percentage of time that a gradient is applied at a specific amplitude?
Duty cycle

*If a gradient is at its resting state 60% of the time during a measurment, it is said to have a 40% duty cycle.
The slew rate is:
The rate of change in gradient strength during a transition from one amplitude to another.
The MRI system's host computer is responsible for:
Communicating the sequence parameters to the rest of the system components.
The actual timing of the gradients is achieved by:
The pulse system controller
The MR signal is digitized by the:
Analog to digital converter or ADC
The Fourier Transform is performed by which MRI system component?
Array processor

* Image reconstruction is possible with the application of a Fourier Transform calculation. Since such calculations are very sophisticated and many thousands of them are required in order to reconstruct a single image, specialized hardware is required which is dedicated to achieving such calculations at high speed.
Which of the following functions is a responsibility of the RF coils?
Detection of the MR signal
What is responsible for the amplification of the MR signal?
Preamplifiers

*It is considered to be a separate component from the RF coils.
Which of the following components is NOT located within the bore of the magnet?
Passive magnetic shielding
The MRI system's array processor is responsible for:
The calculation of the MR signals into the resulting MR image.
The echoes emitted from the patient's body are passed through a pre-amplifier to:
Greatly increase the strength of their tiny signals.
*Uses to ADC
A pulse sequence to select a 3 mm sagittal slice requires 40 amps of electrical current through the gradient coil. If the current is increased to 80 amps, a ________ will result.
Thinner slice
___________ magnets cannot lose their magnetic field without being disassembled.
Permanent
Which of the following is an example of Faraday's law of induction?
Magnetic variations as a result of relaxation mechanisms causing the induction of current into a RF coil.
Faraday's law of induction states that "any magnetic change which occures near a neighboring conductive material will cause a voltage to be induced across that conductive material."
________ Shielding is another name for Faraday shielding.
RF
What types of sheilds control the vaious types of magnetic fields?
Magnetic, gradient, and active sheilding

*They do not function as Faraday shields and do not affect RF transmissions or signals.
Which of the following statements is FALSE regarding the best results obtainable with a receiver coil?
The largest possible receiver coil should always be selected.

*The coil should NOT be much larger that the ROI.
When a different RF coil is used for transmitting than the RF coil used for transmitting than the RF used for receiving, they must be decoupled. What is decoupling?
Electrically isolating the two coils so that the RF energy is only applied to the transmitter coil.
_____________ magnets are routinely used in producing a magnetic field of 1.5T.
Superconducting
_____________ magnets require a constant connection to a power source in order to maintain their magnetic field.
Resistive
Which of the following statements is TRUE?

A. Permanent magnets use more cryogen than superconducting magnets.
B. Cryogens may be refilled while a patient is in the magnet.
C. Leaving the room door open in a room whose walls comprise the Faraday shield may adversely affect the image quality.
D. The use of magnetic shielding allows people with pacemakers to have an MRI exam.
Leaving the room door open in a room whose walls comprise the Faraday shield may adversely affect the image quality.
Which of the following types of magnets can have their magnets can have their magnetic reduced to zero in the shortest amount of time?
Resistive (Can be reduced to zero in a few seconds)
*Superconducting requires a few minutes
A sequence is run to select a transverse slice 5 mm from isocenter, towards the patient's feet, along the z-direction. If the direction of current in the x-gradient coils is reversed:
A mirror image of the same slice would be generated. Reversing the current in the x- gradient will result in left and right being reversed in the image.
___________ magnets are used on most MRI systems produced today.
Superconductive
If gradients takes 1.5 msec to reach an amplitude of 8mT/m, the slew rate is _______ mT/m/sec.
5333

(8 mT/m) / (0.0015 sec)
The fringe field of a 1.5 T superconductive magnet extends along which of the following directions?
all directions
How will an extra steel I-beam in the floor, about 150 cm from the magnet's isocenter, affect the MR system?
It will distort the homogeneity of the main magnetic field.
Which of the following statements is True?
A. Superconducting magnets can produce magnetic fields up to 10 T.
B. Resistive magnets may only have horizontal magnetic fields.
C. Superconducting magnets are never used is MRI systems.
D. Permanent magnets are ideal for use in mobile systems.
A. Superconducting magnets can produce magnetic fields up to 10 T.
A pulse sequence is run to select a transverse slice located 5 mm from isocenter toward the patient's feet, along the z-direction. What would be the result if the direction of current flow in the z-gradient coils were reversed?
A transverse image from tissues located 5 mm from isocenter towards the patient's head, along the z-direction would be generated.
Some manufactures have incorporated a Faraday shield into the design of their magnets, thus eliminating the need to enclose the entire magnet room with RF shielding. Which of the following is TRUE regarding this self shielded type of MR system?
Installation costs will be decreased since expensive metal in the magnet room walls can be avoided.

*Faraday shielding does NOT necessisarily incorporate magnetic shielding, therefore Faraday shielding does NOT allow the trailer to be parked any closer to heavy moving objects.
The actual size of the Faraday room if properly designed should not affect the signal to noise of the images. T/F?
True. It should NOT
A given MRI system has a maximum gradient amplitude of 10 mT/m. For a specific pulse sequence and set of parameters, the prescription of a 3 mm sagittal slice requires an x-gradient amplitude of 8 mT/m, at the narrowest bandwidth. If all parameters remain constant except for slice thickness, a _____ mm sagittal slice is NOT possible on this system.
2

*Compared to the 3 mm slice, a 2 mm slice requires an amplitude of (3mm/ 2 mm) x 8 mT/m =12 mT/m
A region of interest extends about 5 cm below the surface of the skin. If you could only have four circular coils to chose from, a ___ cm diameter coil will have the best sensitivity to signal from the region of interest and at the same time provide the highest signal to noise ratio.
10

Which of the following would result if a 5 cm diameter circular coil was used to image the shoulder of a professional weight lifter?
The portion of the shoulder closest to the coil would have plenty of signal intensity, but the signal intensity from deeper structures would be greatly decreased.

*Images created by circular coils always show the brightest signal adjacent to the coil and a gradual loss of signal in tissues as distant increases from the coil.
Upon installing a new MRI system, the service engineer mistakenly connects the x-gradient amplifier to the y gradient coil. The y gradient coil and the z gradient amplifier to the x gradient coil. What type of image would result when the technologist runs the first sagittal slice?
A coronal image (the machine mistakes the x for a y which is Really a coronal image)
Around which of the following magnets is it necessary to restrict ferromagnetic objects?
All types of magnets