Selection of the incorrect body part from the examination menu may result in _____
|
image misinterpretation
|
|
kilovoltage peak should be chosen ____
|
for the type and amount of contrast desired
|
|
the k-edge of phosphor imaging plates ranges from ____ keV
|
30 to 50
|
|
the range of kilovoltage peak that can be used with CR is ____ KVp
|
45 to 120
|
|
True or False: less kilovoltage peak is needed with conventional radiographic screens than is necessary for CR systems
|
False
|
|
mas is selected for the _____
|
number of electrons needed
|
|
when insufficient light is produced by the imaging plate phosphor, the image will _____
|
be grainy
|
|
quantum mottle is caused by ____
|
insufficient light
|
|
True or False: backscatter from the cassette-detector influences the amount of mAs necessary to produce the image
|
True
|
|
using an imaging plate that is much larger than the part being examined will ______ resolution
|
decrease
|
|
True or False: using the wrong imaging plate size may result in a matrix with very small pixels
|
False
|
|
the higher the number of pixels in a matrix, the _______ the image resolution
|
higher
|
|
true or False: grid lines projected onto the imaging plate will not interfere with the image
|
False
|
|
the moire artifact results when grid lines and scanning lasers are ______
|
parallel
|
|
Grid use for CR examinations is more critical than in film/screen radiography because CR _____
|
imaging plates record more densities
|
|
the number of grid lines per centimeter or lines per inch describes grid ____
|
frequency
|
|
Grid ratio describes the ____
|
relationship between the height of the lead strips to the interspace
|
|
which of the following requires more critical beam centering?
-parallel grids -focused grids -low-ration grids -low-frequency girds |
focused grids
|
|
collimation to a smaller area _____
|
decreases compton interactions
|
|
true or False: it is acceptable to replace collimation with shuttering
|
false
|
|
True or False: images should be marked only with computerized markers
|
False
|
|
the exposure indicator number for Fuji systems is the ____ number
|
S
|
|
the exposure indicator number is an _____
|
indication of radiation absorbed by the imaging plate
|
|
true or False: with Fuji exposure indicators, the lower the S number, the higher the exposure
|
true
|
|
What manufacturer uses exposure indicator numbers in the thousands?
|
Kodak
|
|
As kodak exposure indicator numbers increase by 300, exposure
|
increases by a factor of 2
|
|
an S number of 400 in a Fuji system ____ an S number of 2000
|
is half the exposure of
|
|
in a Fuji system, each change of exposure number of 200 is equal to a change by a factor of ___
|
2
|
|
in a ll systems, exposures outside the range indicate ____
|
both overexposure and underexposure
|
|
a histogram is____
|
a graphic representation of x-ray exposure
|
|
dark line artifacts can be caused by ____
|
scatter through the back of the cassette
|
|
horizontal white lines may be caused by ____
|
dirt on the light guide in the reader
|
|
to avoid the moire grid artifact, the grid must be oriented ____ to the plate readers laser scan direction
|
perpendicular
|
|
fine white lines appearing on a hard copy of the image may be caused by _____
|
debris on the mirror in the laser printer
|