  • Front
  • Back

What is AAC

Artificial Airway Compensation

What must you enter for AAC

size and length of tube

When doesn't AAC work on the Avea

not on mandatory breaths in neonatal PSIMV unless flow cycle is engaged

What is flow cycle on the Avea

cycles to exhalation with flow, not time. Set on 10% means 90% of flow and ventilator will cycle to exhalation

What is rise time

speed of breath delivery 1= fastest, 9= slowest

Where do you set rise time if air hungry?

1 (fastest)



Where do you set rise time for alveolar recruitment

9 (slowest)



What is machine volume on the Avea

delivers a set volume, limited by pressure

What is volume limit on Avea

Limits the volume delivered

What is APRV on the Avea

Airway Pressure Release Ventilation

What do you set for APRV

Th, Tl, Ph,Pl

Is APRV biphasic in the Avea

Yes, but not available in neonatal mode

What is C20

comparison of last 20%, C20/C



What does C20 show

over 1 = no dissension, under 1 = overdistension, beak on pressure/volume loop

How do you determine tube depth

6 + weight in Kg

How do you accurately determine tube size in under 500g baby

X-Ray only

What is a left shift on a pressure/volume loop

stiffer lungs

What is a right shift on a pressure/volume loop

more compliant lungs

What is a jagged flow/pressure loop at the top

water in tubing, inspiratory side

What is a jagged flow/pressure loop at the bottom

airway congestion, expiratory side

What is a figure "8" in the flow/pressure loop

patient pulling flow, increase pressure support

What does a circle mean on the volume, X axis

leak

What is neonatal tube size

2.5 <1000g, 3- 1000-2000g, 4 - 2000-3000g