The greatest enemy of recorded detail is __________.

Motion

SID has its greatest impact on what two image quality features?

Image density


Recorded detail/spatial resolution

Focal spot size impacts what image quality?

Recorded detail/spatial resolution

The features/factors Illustrated here have a significant impact on what quality of the x-ray image?

Scatter

How does each of these features/factors affect image production?

As kV increases, scatter increases


As field size increases, scatter increases


As tissue thickness/density increases, scatter increases

From the following list, select the factors that are most likely to have an impact on image contrast: motion, beam restriction, protective filtration, grids.

Beam restriction


Grids

From the following list, select the factors that are most likely to have an impact on image contrast: part thickness, anode heel effect, pathology.

Part thickness


Pathology

From the following list, select the factors that are most likely to have an impact on image spatial resolution: focal spot size, SID, OID, kVp, mAs.

Focal spot size


SID


OID

From the following list, select the factors that are most likely to have an impact on image spatial resolution: motion, beam restriction, protective filtration, grids.

Motion

From the following list, select the factors that are most likely to have an impact on image spatial resolution: part thickness, anode heel effect, pathology.

Part thickness (resulting in OID)

What exposure error is Illustrated in the image?

Double exposure

Why is this CR image not overexposed?

Because of automatic rescaling

Lack of adequate collimation and/or selection of incorrect processing algorithm is likely to cause __________ errors.

Histogram analysis

From the following list, select the factors that are most likely to have an impact on image distortion: focal spot size, SID, OID, kVp, mAs.

SID


OID

From the following list, select the factors that are most likely to have an impact on image distortion: motion, beam restriction, protected filtration, grids.

Motion

From the following list, select the factors that are most likely to have an impact on image distortion: part thickness, x-ray tube angle, IR angle.

All of them

What was the cause of the exposure problem seen in this image?

Improper positioning of the IP, causing failure of exposure field recognition

What does the abbreviation ROI refer to?

Region of interest

CR artifacts that appear as random small black spots are usually representative of __________.

Fog from environmental sources

What is the name of the computerized system in which the radiographer has a selection of exams available on the console from which to choose?

Anatomically programmed radiography

What is the name of the computerized system in which pre-programmed exposure techniques are available?

Anatomically programmed radiography

What is the name of the device that can be used to measure anatomical parts?

Calipers

What is the name of the exposure system that is used to make small incremental changes in exposure to compensate for variations in body part thickness?

Variable kilovoltage system

What is the name of the exposure system that uses an established, optimum kV for each anatomical part?

Fixed kilovoltage system

What kV increase is indicated for a medium-sized plaster cast?

5 to 8 kV

Using the variable kV exposure system, how is kV adjusted for each centimeter increase or decrease of tissue thickness?

+-2 kV

Select the conditions that are likely to require an increase in exposure: atelectasis, gout, pneumonia, emphysema, osteochondroma.

Atelectasis


Pneumonia


Osteochondroma

Select the conditions that are likely to require a decrease in exposure: congestive heart failure, sclerosis, pneumothorax, osteoporosis, cirrhosis.

Pneumothorax


Osteoporosis

What type of cast often requires no change in exposure?

Fiberglass

How should technical factors be adjusted in pediatric radiography to minimize possible motion?

Short exposure time with higher mA

List some examples of immobilization devices, techniques, and equipment used in pediatric radiography.

Pigg-O-Stat


Papoose board


Sheet wrap


Sandbags


Compression band


Stockinette


Tape


Radiolucent sponges

What are the two basic types of artificial contrast media?

Positive (barium) and negative (air)

What are the two types of compound substances used to make positive contrast media?

Iodine and barium

Artificial contrast media are required when subject contrast is __________.

Low

Artificial contrast agents can be introduced into the body in what two ways?

Injection and ingestion

Which type of contrast agent produces increased brightness in tissues?

Positive

What can be used to increase image contrast when radiographic tissues having low subject contrast?

An artificial contrast agent

What are the component parts of the AEC that function to measure the amount of radiation transmitted?

Detectors/cells

What type of AEC uses a photomultiplier tube?

Phototimer

What type of AEC is considered an "entrance-type" device?

Ionization chamber

When using AEC, the radiographer should set the backup time to __________ of the total expected required exposure time.

150%

Regulations require that AECs have a __________ mAs safety override.

600

Why is part positioning and centering so critical when using AEC?

If the part is not accurately centered to the appropriate detector, over or under exposure is likely to occur

If a PA chest was performed with the center detector selected, the resulting image is likely to be __________.

Overexposed

If a lateral lumbar spine were performed with the center detector selected and the spinous processes centered to the sensor, the resulting image is likely to be __________.

Underexposed

What is the term used to describe the shortest exposure time that a particular AEC system can produce?

Minimum response time

From the following list, select the factors that can have a significant impact on receptor exposure: focal spot size, SID, OID, kVp, mAs.

SID


kVp


mAs

From the following list, select the factors that can have a significant impact on receptor exposure: motion, beam restriction, protective filtration, grids.

Beam restriction


Grids

From the following list, select the factors that can have a significant impact on receptor exposure: part thickness, anode heel effect, pathology.

All of them

For the same FOV, arrange the following Matrix sizes in order of increasing spatial resolution: 1024x1024, 512x512, 4096x4096

512x512


1024x1024


4096x4096

The extent/number of exposure intensities and image receptor can accurately detect is termed __________.

Dynamic range

What is the term used to describe a combination of rows and columns of small picture elements?

Matrix

What term is used to describe the volume elements within a matrix?

Voxel

The larger the matrix size, for a given FOV, the __________ the pixel size and the __________ the spatial resolution.

Smaller


Better/greater

A digital image is made up of a 2D array of numbers called a __________.

Matrix

What term is used to describe the picture elements within a matrix?

Pixel

What formula is used to determine pixel size?

FOV ÷ matrix size