• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/120

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

120 Cards in this Set

  • Front
  • Back
Amperometry
-non-equilibrium electrochemistry
-monitor current at constant applied voltage -- at potential which limiting current will occur
Voltammetry
-non-equilibrium electrochemistry
-monitor current at varying applied voltage -- at Eappl
Electrolytic Cells
-apply external voltage to pair of electrodes to force rxn to occur
-Eappl = Ew - Eref - iR
Anodic Current
negative current
Cathodic Current
positive current
Limiting current
-maximum current that can be achieved in given electrolytic cell
-diffusion limiting
Decomposition Potential
-potential at which current is zero
E1/2
-potential at which 1/2 limiting current occurs
-usually close to the Eo value for Ox + ne- <--> Red
if reaction reversible
-however if electron kinetics are slow then E1/2 < Eo -- diffusion dependent
Diffusion in Amperometry
-the closer the Ox species to the electrode the higher the [Red] will be
-increased electrode area and charge also leads to increased [Red]
Mediator
-mediating species can be used instead of normal reaction to measure concentrations due to chemical limitations of reactants
-ie replacing O2 with Ferrocenium derivative in glucose sensors
Three electrode configuration
-in amperometry/voltametry
-used when current flowing through reference electrode would change the potential of that electrode
-in three electrode config no current passes through reference
Matrix Effect
-sensitivity towards analyte in question altered by other species in sample -- other components change the effective electrode area by adhering to electrode surface
-corrected by standard addition
Standard Addition Technique
1. obtain signal Ix
2. spike sample with known amount and obtain new signal (Ix + Is)
3. use ratio of concentrations to intensities to solve for [x]
Polarography
-type of voltammetry
-uses DME (dropping mercury electrode)
-Hg drop functions as electrode - new drop = fresh electrode
-no adsorption of species in sample
Stripping Voltammetry
-most sensitive of all voltammetric techniques
-use thin Hg film over large carbon electrode
1. apply negative voltage to Hg
-solid metals dissolve into Hg
2. stir solution and rest
3. scan positive voltage
-the result is different metal ions stripping out of Hg at different voltages
Electrogravimetry
-based on plating metal onto electrode and weighing the electrode
Coulometry
-measures total current flow over time
-as analyte consumed, conc decreases and limiting current decreases as well
-need sample to be pure for analyte (100% current efficiency) --> limiting factor
Absorption Spectroscopy
-light source shines light at specific λ through sample and measures light that comes out
-absorbance
Emission Spectroscopy
-measures light emitted by sample after addition of heat
-also measures bioluminescence (no heat added)
-no light source
Fluorescence Spectroscopy
-light added at specific wavelength
-light excites sample and emits light of varying λ
-detector set at 90 angle to avoid contamination from light source
-no nephelometry
Nephelometry
-particle scattering by sample
-light is not absorbed it passes through sample (scatters)
-bac sample spec in lipuma lab is like this
Molecular Photon Emission
Emitted light is lower energy (high λ) then light absorbed
Molecular Fluorescence Spec Setup
1. Light Source
- Tungsten = vis
- Deuterium = UV
2. Excitation monochromator
- selects λ1 = λmax absorbance
3. Sample
4. Emission monochromator (90 degree angle)
- selects λ2 = λmax fluorescence
5. Detector
Red Shift
-phenomenon of energy loss from fluorescence
-photons lose energy due to vibrational and non-radiative energy
- λ2 > λ1 in terms of wavelength --> the red shift
- property unique to molecules and acts as a fingerprint due to its specificity
Fluorescence applications
-most are quantitative
-exploit low detection limits
-use quenching
-chemiluminescence
-used in immunoassays (labeled antibody)
Chemiluminescence
-light emits as a side product of a chemical reaction
-no light source needed
Fluorescin
-common fluorescent indicator
-binds AAs and can therefore be used to label proteins
Quenching
-indirect method of fluorescent detection
-O2 is a known quencher
-relax excited molecules without emitting light
ie FRET
FRET
-quenching method that relies on close proximity of quencher to fluorescing molecule
-quencher absorbs at the λ that fluorescing molecule emits
Flow Cytometry
-cells labeled with fluorescent tags
-apparatus setup so one cell passes through detection area at a time at predetermined rate
-can count number of cells by the number of fluorescent peaks -- histogram
-different cells have different emission spectra (diff number of fluorophores per cell, diff types of fluorphores emitting at diff λs)
Optical Atomic Spec
-Two types-
1. Atomic Absorption
2. Atomic Emission
-primarily measure metallic activity (toxic metals)
Atomic Aborption
-flame vaporizes element and light shown through sample
-light passes through to monochromator and then detector
Hollow Cathode Lamp
-used only for AA as a light source
-emits narrow λ range of light - specific for element of choice
-done by placing element in cathode of lamp and using high voltage to ionize gas
-excited ions strike cathode surface emitting light
-monochromator set to same λ
Atomic Emission
-flame heats and vaporizes sample
-light emitted by excited gas phase atoms of sample into monochromator and then detector
Atomization via Flame
-aspirate sample using nebulizer
-sample then heated by flame and atomized
-high temperature works best for emission and low temperature works best for absorption
Boltzmann Distribution
-equation that shows how T effects emission techniques much more than absorption (negligible change in aa)
-lower T results in clearer spectra but less emission
Flames
-T dependant on oxidant used
-Acetylene/O2 is the hottest and used for emission
-Acetylene/Air is the coolest and used for AA
-in AA light shined through the side to maximize 'b' in the Beer-Lambert Law
Graphite Furnace
-second way to atomize atoms -- more efficient than flame
-used exclusively for AA
-uses a graphite tube heated by an external power source
1. sample dried - low T
2. sample ashed - low T
3. sample atomized - high T
-light is shined through tube containing atomized sample
ICP-E(mission)S(spectroscopy)
-uses ionized Ar gas and oscillating magnetic field to cause a hot plasma of ionized gas
-most sensitive for emission
-much hotter than any flame
ICP-M(ass)S(pectroscopy)
-most sensitive of all atomic methods
-ionize a small fraction of sample (using same magnetic coil as in ES)
-these ions are accelerated into a mass spec where ions with different mass/charge ratios are separated and detected
-each element has its own detection limit -- the lower the better
Spectral Interference
due to overlapping bands for emission or absorption
-solved by choosing a different λ
Chemical Interference
due to species in sample that may hinder atomization -- causes false low value
-solved by using releasing agent to protect atom of interest or sequester species responsible for interference
Ionization Interference
due to other ionizable species in sample releasing electrons into sample and causing error
Glycosylated Hemoglobin
-test for diabetes
-reference method based on cation exchange separation
-glucose permeates RBC if in blood plasma for long time -- glycosylates hemoglobin
Raman Spectroscopy
-excite molecules with powerful external radiation source -- can us any λ as long as it doesn't correspond to electronic transition
-gives information about vibrational energy levels of molecules
-few photons emitted due to low energies - therefore intense light source needed (laser)
-Raman spectra can be used to report on tissue quality -- healthy v. damaged tissue
Chromatography Basics
-separation of solutes through a column containing two miscible phases: mobile and stationary phase
-Based on interactions between mobile phase solutes and stationary phase species
-More interactions = slower movement = later signal
Adsorption Chromatography
-solute adsorb onto surface of stationary phase
-inert particles
Partition Chromatography
-polymer or liquid support coating of solid stationary phase provides binding for solutes in mobile phase
-coated particles
Hybrid Adsorption/Partition Chromatography
-adsorption chromatography with thin monolayer
-polar particle with hydrophobic monolayer -- reversed phase packing materials
-polar molecules wash out first (don't interact with monolayer)
Ion-Exchange Chromatography
-charged functional groups adhered to solid stationary phase bind to ions in solution
-particles coated with charged molecules
-electrostatic interactions
Molecular Exclusion Chromatography
-stationary phase contains pores where smaller molecules can enter
-porous particles
-smaller molecules come out last
Affinity Chromatography
-stationary phase contains molecules with selective interactions for analyte
-particles coated with selective proteins (aka antibody)
-immunoassays
Chromatogram
-detector signal (y) v. time (x)
-peaks represent one molecule each -- peaks should appear gaussian
Tm
- void time
- time for mobile phase to pass through column
-called "air peak" or "dead time" in GC
Tr
- time for solute to exit column
Adjusted Retention Time
Tr-Tm
Equation of Chromatography
Vr = Vm + (k)Vs
-Vm = volume of mobile phase
-Vs = volume of stationary phase
-Vr = retention volume

Vr = F(Tr)... F=flow rate
Vm = F(Tm)
Resolution
-higher = better
-dependant on large peak separation and small band widths
Theoretical Plates
-concept used to describe column effectiveness
-a theoretical plate is essentially one unit of equilibration between mobile phase solute and stationary phase -- imaginary sections of column
-the greater the number of theoretical plates (N) the better the column
Longitudinal Diffusion
-band spreading
-important in GC because gas diffuses much quicker than liquid
-greater diffusion means wider bands -- less efficient column
-decreases with higher flow rate (less time in column)
Mass Transfer Kinetics
-how quick the solute interacts with the stationary phase
-dependant on film thickness, particle size, and velocity of mobile phase
-a higher value means a wider band - mass transfer lags behind mobile phase speed
-faster mobile phase, bigger particles, and thicker film means a wider band
Eddy Diffusion
-depends on the path the solute takes through the packing column
-only found in particle columns
-larger particles = greater diversity of paths = higher diffusion
Mobile phase speed (u) v. Plate Height (h)
-at really low u, h is high (longitudinal diffusion)
-at really high u, h increases (finite mass transfer)
-find the "sweet spot" of intermediate u to minimize h
-determined experimentally
Capillary Columns
-aka 'open tubular'
-no particles (packing) hence no Eddy diffusion
-consists of extremely long tube with very small radius
-stationary phase coats the inside of the tubing
-greater length = greater N
-low back pressure
-small sample size due to small tubing radius
Capillary Column v Packed Column
-capillary chromatogram has more and higher peaks than a packed column
-has lower A and C term than packed column
-has lower back pressure than packed column
Overloading
-[solute] is too high
-peak is non-gaussian -- rear of peak abruptly dips to baseline
Tailing
-stationary phase contains some sites with increased interaction with solute
-peak is non-gaussian -- rear flares out and is wider than gaussian peak
GC mobile phase
-inert gas usually H2 or N2
GC stationary phase
-solid/liquid
-packed or capillary
Packed GC
- GS -- packed column particles
- GL -- liquid/polymer coated particles
Capilary GC
-glass column wall or hydrocarbon solvent covalently bound to wall
GC overview
-sample must be volatile
1. small needle containing volatile liquid or gas injected into injector
2. constant gas flow pushes sample into column
3. detector determines elution and represents as peak
Injection
-poor reproducibility (usually use internal standard technique)
-1-10 ul volumes used
-injector chamber preheated to temp >50
Polydimethylsiloxane
-all-purpose GC stationary phase
-non-polar coating added to quartz column
-steps
1. coat with DMCS
2. add alcohol
3. add more DMCS to create film
-substitute methyls with phenyls to make column more non-polar
Polyethylene Glycol
-polar GC stationary phase
-great for acids and alcohols
Polydiphenylsiloxane
-more non-polar stationary phase of GC column
Trifluoroporpyl siloxane coating
-non-polar GC coating
GC Elution
-based on boiling point (vapor pressure) and solvation (miscibility) of analytes
-lower b.p. tends to elute first
-polarity however may cause species to elute much later
GC Temperature
-too low = really long experiment times
-too high = poor peak resolution
-linear temperature ramp is a solution -- linearly raise temperature to both shorten time and maintain resolution
Thermal Conductivity Detector
-sensitive to inorganics and hydrocarbons
-carrier gas is H2 or He
-as species elute the temp of the wire changes -- wire resistance changes
-needs calibration
-0.3 ng detection limit
Flame Ionization Detector
-most common detector
-good for hydrocarbons
-analyte burned in flame
-ions produced are detected as current on collector electrode
-4 pg detection limit
Electron Capture Detector
-uses beta particle decay to ionize carrier gas
-free electrons ionize analyte which registers as current drop
-detection limit in fg
GC and LC Mass Spec
-GC - leak small amt of carrier gas into mass spec
-LC - use electrospray ionization source
-gain retention time elution as well as mass spec data for one sample -- tool to further show species presence in sample
HPLC
-high pressure liquid chromatography
-use very small particles
-high pressure needed to pass solution through limited void space
-small void space ensures mass transfer
HPLC packing
-porous silica particle
-surface modified with monolayers (can be non-polar)
Normal Phase LC
-polar stationary phase and non-polar mobile phase
-non-polar elute first
Reversed Phase LC
-non-polar stationary phase (silica coated with non-polar monolayer)
-polar mobile phase
-polar elute first
Mobile v Stationary Phase
-mp alters sp structure based on polarity
-polarity of mp controls LC peak separation -- like temp in GC
Particle Size
-very important in LC
-smaller particle size = smaller C term = less slope = greater N = greater resolution
Mobile Phase Reservoir
-HPLC component
-attached to pump
Pump
-uses high pressure to force liquid through small particle column
-pushes liquid into injection valve
Injection Valve
-injects reproducible volume of mp into column
1. loads sample into loop in excess (loop filled)
2. valve links MP and pushes entire loop volume into column
-uses 25-100 ul volume
Guard Column
-small length of column to protect actual column from species in sample which may irreversibly bind SP
UV-Vis Absorption Detector
-most common
-light source shines light into eluant (monochromator may be used)
-coupled with diode array to obtain full spectrum of each solute
-detection limit: 1 uM - 1nM (solute dependent)
-MP must not absorb strongly
Fluorescence Detectors
-excite sample with xenon lamp or laser
-monochromator used before or after sample
-detection limit: 1 pM- 1 nM
Electrochemical Detector
-amperometric most common
-three electrodes (reference, counter, and working)
-detection limit: 0.1 to 1 nM
-used to detect neurotransmitters and "native" electrochemistry
Ion Exchange LC
-HPLC method
-SP is charged and MP contains ions - LC based on these interactions
-larger ions and more charged ions are retained longer (ionic radius is what matters)
-use polystyrene-based columns (not silica)
Cation Exchange Column
-requires fixed concentration of given cation in MP - usually by using acidic pH of mobile phase
-SP usually sulfonated
Suppressor Column
-in ion exchange LC
-added column after separation column to neutralize charged MP species and limit background conductivity
-sample ions measured at low conductivity
Ion-Pairing Chromatography
-uses regular reversed phase HPLC column
-add surfactant to MP to coat SP with ion-exchange sites
-used to separate organic ions
LC MP Composition
-low conc = long elution times
-high conc = poor resolution
-use gradient elution to solve problem
Gradient Elution
-changing MP composition during separation (increasing strength) can decrease elution times but maintain resolution
-in ion-exchange LC increasing the MP ion concentration will have the same effect
Molecular Exclusion Chromatography
-used to separate and purify proteins and polymers
-separation based on size/shape/MW
-porous silica or polystyrene particles 'trap' smaller species
-smallest come out last
Affinity Chromatography
-used to isolate or pre-concentrate one specific compound
-SP possesses bound molecules with affinity for compound -- undesired solutes washed off column
-change in pH or ionic strength of MP allows elution of compound to then occur
Electrophoresis
-separation by migration in an electric field
-based on size and charge
Electrophoretic Mobility
μ = zF/f... F = faradays constant, f = hydrodynamic radius
U = (μ)(E)... E = electric field, U = velocity (cm/s)
PAGE
-polyacrylamide gel electrophoresis
-polyacrylamide = cross-linked polymer -- sieves molecules using size
-used for protein (+SDS) and DNA
Coomasie Blue
Protein PAGE stain
Albumin
-2nd highest concentrated protein in blood (hemoglobin first)
-maintains blood vessel integrity and can also act as a carrier protein
Alpha-1-globulins
-high-density lipoprotein (HDL)
-"good" cholesterol
Alpha-2-globulins
-includes haptoglobin - binds hemoglobin
Beta-globulin proteins
-carrier proteins (ie Fe)
-help fight infection
Gamma Globulin
-antibodies
MGUS
-autoimmune disorder
-predominantly found in men
-patients produce very high concentrations of antibody with unknown fxn
-possible precursor to cancer (multiple myeloma)
SDS
-sodium dodecyl sulfate
-surfactant that adds -ve charge to protein peptide backbone
-gives all protein same mobility in electric field
Capillary Electrophoresis
-allows for much higher voltages due to excellent heat dissipation (low current)
-therefore faster separations
-silica capillary flows from +ve to -ve
-dangerously high voltages used (100-300 V/cm)
-no A or C terms due to electroosmotic flow
Electroosmotic flow
-flow due to electrochemical properties
-consistent throughout tubing - unlike pump flow (laminar flow)
-higher number of theoretical plates
Electrophoresis Migration Order
- Cations first -- higher charge and smaller size first of these
- Neutrals second -- all migrate together
- Anions last -- higher charge and smaller size come last
Capillary Electrophoresis Detection Limits
UV-Vis = 1 uM or 1 fmol
Fluorescence = 1 pM - 1 uM
Electrochemical = 1-10 nM
MS = 0.1 to 1 uM
MEKC
-Micellar Electrokinetic Chromatography
-Negatively charged surfactant (SDS) used to form micelles with neutrally charged species in center
-Separates out neutrally charged species based on lipophilicity -- the more micelles bound the more negative the complex and the later it elutes
-Used for separating out steroids