Image of Artifact

Modality


Name of cause

CR


Double Exposure


Extraneous materials blocking image

Upside down grid, loss of density away from center of grid

Grid cut off

Moire pattern, two grids

Motion

Motion with single phase generator

Backscatter from film tray

Over exposed

Underexposed

Film Radiography


Finger nail marks

Film Radiography


Pressure artifact crushed by other films in the bin

Film Radiography


Static: spot or smudge


From ionizing gas molecules near film surface

Film Radiography


Static: tree like


From non-light producing static discharges

Film Radiography


Safelight fog



Film Radiography


Light fog from a cracked door

Film Radiography


Light fog

Film Radiography


Dirt screen

Film Radiography


Poor Film-Screen Contact

Quantum Mottle

Overexposure causing saturation and loss of image in the thinner areas of the patient

Planking

Fading, light image of decreased quality which appears grainy


Information loss is non-uniform

Light leak

CR


Dirty light guide

Faulty Transfer

Faulty transfer

Misregistration

Border detection

Moire

Überschwinger

Density Threshold

US


Reverberation

US


Twinkling

CR


double exposure

CR


Paradoxic Overexposure

CR


Light Leak

CR


Dirty Light Guide

CT


Ring artifact Bad Detector

CT


Beam Hardening

CT


Stair step

CT


Out of Field

CT


Attenuation Artifact

MR


Chemical Shift

MR


Magnetic Susceptibility

MR


Cross Talk

MR


Wrap around

US


Speed Propagation

US


Electrical interference

US


Slice Thickness Pseudosludge

US


Reverberation

NM


Inadequate Counts

NM


Septal Penetration

NM


Camera Contamination

NM


Malfunctioning PMT


CT


Aliasing from a high frequency object



CT


Motion



CT


Windmill artifact


Interpolation from multislice CT



CT


Cone Beam Truncation



CT


Step Artifact


Slice is too thick for reformat



CT


Zebra Artifact


Helical data interpretation



CT


Volume averaging


From slice thickness, produces pseudo-lesions



CT


Out of field artifact


Anatomy is outside the FOV



CT


Algorithm Rebound

CT


Ring artifact


Faulty detector



CT


Beam Hardening



CT


Streak artifact


From bone or contrast



US


Edge Shadowing



US


Distal Acoustic Enhancement



US


Refraction


Duplication of structure



US


Reverberation


Comet tail; tapered triangle; ring down



US


Slice thickness


pseudo sludge (curved); resembles side lobe artifact



US


Mirror image


Usually GB and diaphragm



US


Speed propagation artifact

US


Electrical interference

DR


Planking


Also shows memory artifact



DR


Calibration mask artifact

CR


Cassette upside down

CR


Cassette upside down



CR


Double Exposure

DR


Quantum Mottle


Increase mAs to fix

DR


Paradoxic overexposure

CR


Light Leak

CR


Delayed reading


Grainy image

CR


dirt on plate

CR


Dirty light guide

DR


Moire artifact


Grid line/sampling artifact

DR


Faulty Data transfer

CR


Auto shutter error

DR


Incorrect anatomic region selected

NM


Cracked NaI crystal

NM


Attenuated intrinsic flood source


Metal needle attenuates portion of Tc source

NM


Malfunctioning PM tube


Hot or cold areas on the flood


Fix: retune

NM


Blown PM tube

NM


Edge packing


Reflecting of light at edge of crystal so counted twice by PMT


Intense ring at edge

NM


Septal penetration


High energy photon with low energy collimator

NM


Collimator Septal damage

NM


Pixel roll over


Bit depth is too low

NM


Digital round off error


elephant skin flood

NM


Off peak flood

NM


Uncorrected and corrected flood

NM


Non-linearity


Out of tune PMT


Pincushion, not straight lines on the phantom

NM


Air bubble in flood

NM


Improperly mixed water phantom

NM


Contamination on patient

NM


Metallic artifact

NM


Glandular uptake of free Tc

NM


Poor anatomic detail


Edema, metabolic, increased pt -camera distance

NM


Bladder hot spot


Use shielding

NM


Injection site hot spot

NM


Hot source near camera during acquisition

NM


Moire artifact


Sample too low

MR


Ghosting


Motion in the phase encoding direction

MRI


Aliasing/wrap around


Anatomy outside of FOV

MR


Chemical shift


Occurs in frequency direction, use lower field strength, incr bandwidth, swap phase/frequency

MR


Chemical misregistration

MR


Truncation


Banding in the phase direction

MR


Crosstalk


Need 30% gap between slices and no overlap of slabs

MR


Zipper artifact


RF leak or object in room, type of electrical interference