kVp range used in mammography
|
20 - 35 kVp
|
|
Two advantages of using low kVp in mammography are:
|
Producing high radiographic contrast
Producing a very low energy x-ray beam |
|
The anode heel effect can be used in mammography by positioning the ______ toward the ______ wall
|
cathode
chest |
|
Breast tissue is made up of what 3 tissues?
|
Adipose tissue
Fibrous tissue Glandular tissue |
|
Anodes used in mammography are typically made of what?
|
Molybdenum
|
|
The exit window of the x-ray tube in mammography can be made of what two materials?
|
Beryllium and borosilicate glass
|
|
Magnification projections of the breast provide what three things?
|
Increased resolution
Reduction in scatter radiation Improved visibility of detail |
|
The use of a grid in mammography does what three things?
|
Improves radiographic quality
Increases patient dose Reduces scatter reaching the image receptor |
|
Breast compression will reduce ______
|
magnification
|
|
Breast tissue has very _____ inherent subject contrast
|
low
|
|
AEC's should not be used for mammography (true/fasle)
|
False
|
|
Magnification radiography of the breast ______ patient exposure
|
increases
|
|
What is the focal spot size range used in mammography?
|
0.1 - 0.6 mm
|
|
When magnification mammography is used, the magnification factor used is in what range?
|
1.5 - 2
|
|
Molybdenum is used in mammography x-ray tube targets to create useful _______ characteristic x-rays
|
characteristic
|
|
The low kVp used in mammography will increase what ?
|
photoelectric absorption
|
|
Quality mammograms can't be produced at high 90 kVp because differential absorption is too ____
|
low
|
|
Compared to a general diagnostic x-ray unit, the half value layer of a mammography x-ray unit is very _____
|
low
|
|
What is the minimal total tube filtration of a mammography x-ray unit?
|
0.5 mm Al equivalent
|
|
The increased photoelectric interactions from using a low kVp will create ____ contrast, and ______ patient dose
|
higher
higher |
|
Dedicated mammography x-ray units use what type of generator?
|
High frequency (1% ripple factor)
|
|
The majority of the beam is made up of what type of radiation?
|
Characteristic
|
|
What energy range of photons will maximize subject contrast, and allow the visualization of microcalcifications?
|
17-20 keV
|
|
If the energy levels of the beam are too high:
1. Breast tissue will be ____ penetrated 2. produce _____ scatter 3. ______ image contrast |
1. over
2. more 3. decrease |
|
A smaller focal spot will ______ recorded detail
|
increase
|
|
Beryllium is used as in the port window to allow _______ energy photons to exit the tube.
|
low
|
|
Two geometric problems to overcome are:
|
1. Decreased resolution at chest wall
2. Missed tissue along chest wall |
|
The geometric problems are compensated by what 3 things?
|
1. Heel effect
2. Off center placement of the tube 3. Elimination of cathode side of the beam |
|
The heel effect is very pronounced due to what?
|
Short SID
|
|
Tube configuration uses effect to an advantage by placing the _______ toward the chest wall and the _____ toward the nipple
|
cathode
anode |
|
The ______ axis is used to line up the tube
|
reference
|
|
The effective target angle of the anode is:
|
22-24 degrees
(more than general x-ray due to SID, off centering, and anode heel effect) |
|
Compression of the breast (25-45 lbs) will do what 3 things?
|
1. Improve spatial resolution (see more anatomical tissue)
2. Improve contrast 3. Lower patient dose |
|
Mammography screen film systems use ______ emulsion film and _____ intensifying screen
|
single
one |
|
Selenuim direct capture detector has a direct converesion of x-ray energy to what?
|
Electrical signal
|
|
1 in __ woman will develop brest cancer
|
8
|
|
At low energy _____ predominates over compton scattering
|
photoelectric absorption
|
|
low kvp must be used to maximize _________ effect
|
photoelectric
|