Dissolved Gas Analysis Essay

Great Essays
1.0 REFERENCES
1.1 Sample Preparation and Calculations for Dissolved Gas Analysis in Water Samples Using a GC Headspace Equilibration Technique, EPA National Risk Management Research Laboratory, RSKSOP-175, Rev.2, May 2004.
1.2 Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID), Pennsylvania Department of Environmental Protection Bureau of Laboratories, PA-DEP 3686, Rev.0, April 2012.
1.3 Lange’s Handbook of Chemistry, 14th edition (1992), McGraw Hill.
1.4 Gas Encyclopedia, Air Liquide, online edition (2009).
2.0 SCOPE AND APPLICATION
2.1 Samples reported using this procedure reference the sample preparation SOP commonly referred to as “RSK-175” for use in West Virginia. The
…show more content…
Analytes are detected using a Flame Ionization Detector (FID). The amount of gas in the original liquid sample is directly proportional to the amount of gas in the headspace of a closed vessel.
3.2 Standards are prepared by saturating a solution of water with the gases, and making dilutions from that stock.
4.0 INTERFERENCES
4.1 Since these compounds are highly volatile, care must be taken during sampling to prevent agitation, creation of bubbles or headspace during storage. Transfer all samples quickly to the sampler vials while still cold, if possible.
4.2 Results are dependant on the headspace sampler conditions remaining constant.
4.3 Studies have not been done to determine the effect of differing matrices (chlorination, salinity) on solubility of the gases in this procedure. It should have little effect as the gases have a low partition coefficient.
4.4 Methane is a common contaminate and occurs naturally in the atmosphere. Analysis of a blank proves that the system is not
…show more content…
Each gas is prepared and calibrated separately.
8.3 The chromatography software performs all calculations for determining the calibration model (or fit). Average calibration factor represents the slope of the line between the response for each standard and the origin.
8.3.1 The variation over the range of the calibration must be less than or equal to 20% Relative Standard Deviation (RSD) to be considered linear.
8.3.2 Linear calibration using a least squares regression can also be used. The “goodness of fit” of the linear regression, given as r2 by the software, must be greater than or equal to 0.995.
8.4 Verify the calibration curve with an independent standard for each gas. The second source ICV must be within 20% of its expected value for each analyte.
8.5 Also examine the calculated value of the lowest calibration standard (MRL). This step is used to ensure analytes can be accurately reported at the quantitation limit or to demonstrate bias in the calibration model. If the value exceeds 50% of the expected value, inspect the curve for proper values, proper integration and identification, or ensure proper standards were

Related Documents

  • Improved Essays

    Organization Bunge Nipawin, is a canola crushing and refining plant in Nipawin, Saskatchewan, which is part of a larger company called Bunge Limited. Bunge Nipawin specifically sells canola meal, crude canola oil, and refined canola oil to various external and internal customers. As the refinery and utilities supervisor, I look after the tank farm inventory, the bleaching process, the deodorizing process, the boilers and auxiliary equipment, the tailings pond, and the skimming pits. The Pitch - Problem Background It is industry practice to hinge a quality specification on a taste test, prior to shipping the rail car of canola oil to the customer.…

    • 939 Words
    • 4 Pages
    Improved Essays
  • Superior Essays

    Pah Lab

    • 1911 Words
    • 8 Pages

    Determination of Most Advantageous Sequestering Agent By: Kayla Jeeter and Ashley Driscoll Chemistry 112-508 Abstract The creation and execution of this project served to determine what material is a better sequestering agent of the polycyclic aromatic hydrocarbons (PAHs) that are found in Texas’ water. Charcoal is currently the most typical agent used to sequester PAHs.…

    • 1911 Words
    • 8 Pages
    Superior Essays
  • Improved Essays

    Nt1310 Unit 5 Lab Report

    • 695 Words
    • 3 Pages

    CORRECTIVE ACTION PLAN The laboratory’s proposed action and activities already taken to the noted deficiencies are noted below. Item 1. Chain of custody (COC) G1505428 did not list the requested parameters for analysis. The sample had been in house for several weeks and had not been through the laboratory's second level review process at the time of the inspection.…

    • 695 Words
    • 3 Pages
    Improved Essays
  • Decent Essays

    Nt1310 Unit 5

    • 656 Words
    • 3 Pages

    1 University of the People Unit 5 Written Assignment Analyze the report and present a summary, in your own words, of the failures and your comments on them. A preliminary investigation of flight data showed: nominal behavior of the launcher up to H0 36 seconds; failure of the back-up Inertial Reference System followed immediately by failure of the active Inertial Reference System; swiveling into the extreme position of the nozzles of the two solid boosters and, slightly later, of the Vulcain engine, causing the launcher to veer abruptly; self- destruction of the launcher correctly triggered by rupture of the links between the solid boosters and the core stage. The origin of the failure was thus rapidly narrowed down to the flight control system…

    • 656 Words
    • 3 Pages
    Decent Essays
  • Superior Essays

    The warm up of the probe is done through the LoggerLite program under the file metabolism experiment. The chamber needs to be filled with 200 mL fish water while the probe is calibrating. During the probe calibration, weight a separate container with 200 mL fish water on the analytical balance. Then obtain two similar sized goldfish and weight the goldfish with the analytical balance already tare with the weight of the container and 200 mL fish water. Record down the weight of the 2 goldfish.…

    • 1252 Words
    • 5 Pages
    Superior Essays
  • Improved Essays

    Unknown Lab Report

    • 750 Words
    • 3 Pages

    Materials. Numerous substances in the experiment were used. The most frequently used was the unknown due to the need to test its physical and chemical qualities. When a solution of the unknown was made, 1.000 g of the unknown and 1.0 mL of water was used to make it. To test for the possible ions, 1.0 mL of silver nitrate and 1.0 mL of nitric acid were used for the ion test.…

    • 750 Words
    • 3 Pages
    Improved Essays
  • Improved Essays

    Acceleration Lab Report

    • 410 Words
    • 2 Pages

    The definition of an accelerant is a substance that accelerates the speed of fire or makes a fire more intense (Accelerant | Define Accelerant at Dictionary.com, n.d.). When forensic laboratories have debris that was collected from a crime scene they can use either the Headspace or the Vapor Concentration techniques to determine if there was an accelerant used. Both techniques use Gas Chromatography, which is separating mixtures based on their distribution between a stationary liquid phase and moving gas phase (Saferstein, 2011). The Headspace Technique was developed in the 1980’s to aid in explaining the odor compounds found in the air surrounding objects such as plants, flowers and food (Headspace technology - Wikipedia, the free encyclopedia, n.d.) Headspace analysis is the investigation of the components present in the gas, which is found at the top of the vial (Labhut Education Centre | Principles of GC Headspace Analysis, n.d.).…

    • 410 Words
    • 2 Pages
    Improved Essays
  • Great Essays

    Seven trials were done, with three chosen as the most precise (accurate to 4-5 drops). The pipette was then cleaned, placed back in test tube and excess hexanes disposed of into the hydrocarbon waste. Next a retort stand was set up with an iron ring attached with a watch glass placed on top about thirty centimetres from the lab bench. The watch glass was obtained from the fume hood, with tongs and was rinses with tap water and then distilled water. After being placed, the watch glass was filled to the edges with distilled water, and the diameter measured with a ruler ( 0.1cm).…

    • 1814 Words
    • 8 Pages
    Great Essays
  • Improved Essays

    Albuterolfate Case Study

    • 184 Words
    • 1 Pages

    The curve also should include a total of five levels (50%, 75%, 100%, 125%, and 150% of the target concentration “based on the sample preparations”) 2- Specificity: injecting sample , standard , standard spiked with 25% sample. Examine that there…

    • 184 Words
    • 1 Pages
    Improved Essays
  • Great Essays

    Flame Test Lab Conclusion

    • 1167 Words
    • 5 Pages

    The metal ions BsCl and LiCl are in the unknown solutions A and B from the barrels on the vacant lot. 3. A student performed flame tests on several unknown substances and observed that all of the flame colors were shades of red. What could the student do to correctly identify these substances?…

    • 1167 Words
    • 5 Pages
    Great Essays
  • Great Essays

    Drop Time Experiment

    • 2013 Words
    • 9 Pages

    Determining the effect of the mass on the drop time of a parachute Introduction and hypothesis The drop time of a parachute is dependent on multiple factors. Apart from mass, there is the obvious height, area of the parachute and other external factors such as wind. Mass is one of the main effectors of the drop time of a parachute, which can be proven through Newton’s second law, F=ma…

    • 2013 Words
    • 9 Pages
    Great Essays
  • Improved Essays

    Allegro Case Study

    • 816 Words
    • 4 Pages

    1) There is no major Quality issues at Allegro as the meeting mainly review and discuss unit tilted issue faced for package OT1084 (ALGRO 3) & OT1085 (ALGRO 4) and the measurement method. 2) We presented why C-Pak unable to measure the pocket size/top opening according to Allegro measurement method especially for A0, A1 and B0 dimension as below:- 3) In fact, we’ve tried few approach using Micro Vu to measure a plane 0.3mm above the bottom of the pocket however the optical measurement unable to determine the pocket wall point clearly once it’s been lift up 0.3mm. 4) Therefore C-Pak has requested Allegro their existing supplier (Advantek) measurement method for review but it was neglected by Allegro engineering due…

    • 816 Words
    • 4 Pages
    Improved Essays
  • Superior Essays

    Initially this was not a popular technique as there were incompatibility problems between the two instruments. Liquid Chromatography Mass Spectrometry Flow Rate: 1ml/min = 500ml/min gas flow 10ml/min gas flow Liquid Phase Operation Vacuum Operation 25-50oC 200-300oC Table 1: Requirements for both liquid chromatography and mass spectrometry In order to overcome the incompatibility issues, technological interfaces for LC-MS were developed. Without an interface, upon injection of the analyte the liquid would vaporize and large amounts of gas would get into the MS unit. The interface transfers the analyte in solution phase from the LC column to the high vacuum of the MS as a single molecule in the gas phase.…

    • 1044 Words
    • 5 Pages
    Superior Essays
  • Great Essays

    Validation: The validation was performed for fallowing method for these analytical parameters System suitability, Accuracy, Precision, Linearity, LOQ, LOD, Robustness. a) Accuracy The closeness of agreement between the value, which are accepted either as a conventional true value or an accepted reference value and the value found. Accuracy is represented and determined by recovery studies. b) Precision…

    • 1049 Words
    • 5 Pages
    Great Essays
  • Great Essays

    Cyclic Voltammetry of Ferricyanide Analytical Lab 1: 2410L Benjamin A. Battiste Group 2 10/14/15 Introduction: The purpose of this experiment is to use cyclic voltammetry to see how ferricyanide can be reduced to ferrocyanide. Using cyclic voltammetry a calibration curve will be created to determine the concention(mM) of an unknown solution of ferricyanide.…

    • 1061 Words
    • 5 Pages
    Great Essays