Elekta Synergy S linear accelerator (LINAC) was used to produce the 6, 10, and 15MV photon beams for measurements. The accelerator is fitted with the new Agility multileaf collimators (MLC) thinner than the previously used MLC’s to improve a field definition (Kantz et al. 2015; Thompson et al. 2014). The quality assurance was carried out monthly to ensure the LINAC response does not vary from the commissioning data (CPQR 2013). The LINAC has been calibrated to 100 cGy/MU at Dmax for a 10×10 cm2, and daily output checking was performed to ensure that LINAC operated within the ±3% limit (Andreo et al. 2006). 3.2 Film
3.2.1 Kodak X-OMAT V/EDR2 film
3.2.2.1 Calibration curve
The radiographic EDR2 films which were used for investigation of PDD,
…show more content…
The right image shows the sensitive volume of a diode or diamond detector
The detectors were aligned visually with the cross hairs (centre of the beam, CAX), and then the OmniPro software is used to correct the positioning of the detector. The servo measured a depth dose curve to find the position of Dmax and 40 % of a maximum dose. Then in-line and cross-lines profiles are measured at these depths, and the position of the 50% level was determined for these profiles. The positions of the 50% for the profiles are compared and the offsets are then displayed.
3.3.2.2 Voltage
Diode and micro-diamond detectors are semiconductor detectors thus no voltage is applied, CC01 required 300 volts to allow for optimal data
…show more content…
The two detector method employs the use of a field and a reference detector. The reference detector is used to minimize the effect of dose rate fluctuations (Das et al. 2008). The single detector method utilizes the step-by-step mode of data collection, this mode moves the detector at specified step sizes and stops for a specified time. This minimizes the linac dose rate fluctuation which is the same function performed by the reference detector. The pause time was determined by comparing the profiles obtained from two detector method and the single detector method to optimize both the beam on time and the noise.
The single detector mode is further optimized by utilizing two algorithms that are within the Omnipro software. The gradient algorithm is used in measuring beam profiles as the detector’s step sizes alter if two positions have a difference that is more than predicted. The breakpoint algorithm was used for PDD measurements, as the algorithm changed the step sizes after a certain depth was reached. The output factors were measured by collecting signals with the sensitive volume of the detector placed at 10cm

Related Documents

  • Decent Essays

    Unit 3 Physics Lab 3

    • 546 Words
    • 3 Pages

    Further, this experiment sought to clarify the circumstances under which the instruments fail to comply with expectations. Results/Data Part 1: The Motion Sensor Motion Sensor’s Measurement of Distance on ‘Short Range’ 1 m Closest Measurable Distance on ‘Short Range’ 0.12 m Closest Measurable Distance on ‘Long Range’ 0.12 m Distance from Center Line for First Detection of Block on ‘Short Range’ 0.05 m Distance from Center Line for First Detection of Block on ‘Long Range’ 0.055m Table 1:…

    • 546 Words
    • 3 Pages
    Decent Essays
  • Improved Essays

    Distance And Age Of M52

    • 1078 Words
    • 4 Pages

    The upper point on the main sequence which is the most densely populated with stars where the red giants seem to begin is called the turnoff point; the exact location of turnoff point indicates the age of the cluster. Deriving the distance and age of M52 We have already identified main sequence, turnoff point, and giant red for M52 shown in figure by comparing with Figure 1. A best fit line this will help to find the distance to open cluster M52 by using the techniques of best fit line.…

    • 1078 Words
    • 4 Pages
    Improved Essays
  • Improved Essays

    Noxubee Refuge Lab Report

    • 418 Words
    • 2 Pages

    The depth finder was used to determine the depth of the water. The Secchi test determines the visibility depth. The Secchi is a round disk with two white triangles separated by two black triangles. These triangles cover the entire disk. The disk is connected to a tape measure.…

    • 418 Words
    • 2 Pages
    Improved Essays
  • Improved Essays

    Furthermore, guidelines considering the selection of algorithms and input parameters for the MHNSGA algorithm are resulted from the analysis of real-world databases. When it comes to solving the controller placement problem with tens of millions placements for which performing the exhaustive evaluation requires a considerable amount of time and memory budget, our proposed heuristic approach is an appropriate choice. As described before, for such these large-scale instances, it is only possible to calculate an upper bound for evaluated placements and nothing can be expressed about the obtained accuracy of the heuristic algorithm. This is due to not existing the actual Pareto optimal solutions and hence, the absence of reference data to compare.…

    • 404 Words
    • 2 Pages
    Improved Essays
  • Improved Essays

    Acceleration Lab Report

    • 410 Words
    • 2 Pages

    The definition of an accelerant is a substance that accelerates the speed of fire or makes a fire more intense (Accelerant | Define Accelerant at Dictionary.com, n.d.). When forensic laboratories have debris that was collected from a crime scene they can use either the Headspace or the Vapor Concentration techniques to determine if there was an accelerant used. Both techniques use Gas Chromatography, which is separating mixtures based on their distribution between a stationary liquid phase and moving gas phase (Saferstein, 2011). The Headspace Technique was developed in the 1980’s to aid in explaining the odor compounds found in the air surrounding objects such as plants, flowers and food (Headspace technology - Wikipedia, the free encyclopedia, n.d.) Headspace analysis is the investigation of the components present in the gas, which is found at the top of the vial (Labhut Education Centre | Principles of GC Headspace Analysis, n.d.).…

    • 410 Words
    • 2 Pages
    Improved Essays
  • Improved Essays

    Departmental QA Checklist

    • 497 Words
    • 2 Pages

    One camera on the HDR after loader, another where the source enters/exits the patient and the last one monitoring the “radiation monitor” lights inside the room. It is so important to have audio and visual communication during radiation therapy treatment. Especially in HDR treatments, the treating oncology staff, particularly the doctor needs to be able to monitor the patient at all times and visually sees the source smoothly enter and exit the patient. • The survey meter needs to be turned on and tested for optimum operation before treatment • Verification of current activity of radiation source. This is so important because if the calculation of the activity is incorrect, and the patient is treated, then the patient risks being under dosed or overdosed for that treatment due to the miscalculation and incorrect treatment…

    • 497 Words
    • 2 Pages
    Improved Essays
  • Decent Essays

    EXPERIENCE: Amerapex Corporation, Non-Destructive Technician (NDT) 07/2015 - Present NDT level II UT-T & VT/ Level I Guided & UT-SW/ Level I Radiographer/ Safety Representative • Conduct multiple types of non-destructive testing methods on welds, piping and vessels. • Corrosion Under Insulation (CUI) inspections.…

    • 515 Words
    • 3 Pages
    Decent Essays
  • Improved Essays

    Purpose: Our previous computer models suggested that intraluminal thrombus (ILT) within an abdominal aortic aneurysm (AAA) attenuates oxygen diffusion to the AAA wall, possibly causing localized hypoxia and contributing to wall weakening. The purpose of this work was to investigate this possibility. Methods: In one arm of this study, patients with AAA were placed in one of two groups: (1) those with an ILT of 4- mm or greater thickness on the anterior surface or (2) those with little (< 4 mm) or no ILT at this site. During surgical resection but before aortic cross-clamping, a needle-type polarographic partial pressure of oxygen (PO2) electrode was inserted into the wall of the exposed AAA, and the PO2 was measured.…

    • 517 Words
    • 3 Pages
    Improved Essays
  • Improved Essays

    In 1896 Roentgen presented the idea of the x-ray. Only a few months later, systems were already being designed to use the x-ray for diagnosis. Within 3 years, the radiation was being used in cancer treatment. A single idea for treatment grew in a short period of time, with the help of technology. Today we are seeing scientific advances like never before.…

    • 491 Words
    • 2 Pages
    Improved Essays
  • Great Essays

    Place an object at one end of the desk (image A). Using masking tape, label this point C. 4. From the end of the desk opposite the object, estimate the distance to the object. Record your estimate 5.…

    • 976 Words
    • 4 Pages
    Great Essays
  • Improved Essays

    Cole Vickery COM 114 Section 749 Alfa3 Specific Purpose: To inform my audience about the new airport security scanner Alfa3. Thesis Statement: With its advanced scanning process, the new Alfa3 airport security device will not only make airports checks more thorough but also much less of a hassle. INTRODUCTION: I. Attention Getter: Benjamin Franklin said, “If we restrict liberty to attain security we will lose them both.” In the matter present day airport security I believe this applies.…

    • 575 Words
    • 3 Pages
    Improved Essays
  • Improved Essays

    Laboratory Techniques and Measurements Tiffany Abbe Introduction: In this lab I will be adapting more about how to utilize distinctive estimation instruments and comprehend the diverse procedures included. Figuring out how to ascertain temperature and length. Will likewise make estimations utilizing the SI frameworks for fixation, thickness, and volume. While having the capacity to reate arrangements of building so as to change focuses and densities as tock arrangement.…

    • 1009 Words
    • 5 Pages
    Improved Essays
  • Great Essays

    It uses physical markers, or fiducials, which are surgically embedded in and around a tumor to locate it within the body. With the assistance of the fiducials, high doses of radiation can be delivered to the tumor with submillimeter accuracy. In recent years, real time imaging of the fiducials allows SBRT to account for a tumor’s movement caused by breathing. Typical markers are small gold cylinders roughly 0.8 x 5mm.…

    • 1998 Words
    • 8 Pages
    Great Essays
  • Great Essays

    What is radioactive dating? and how does it work? Radioactive, or Radiometric dating is a technique used to determine the age of objects such as rocks or carbon by using the decay rate of radioactive isotopes. It has been used since its invention by Ernest Rutherford in 1905. Radioactive decay itself, is the process by which an unstable atomic nucleus loses energy by emitting radiation.…

    • 1106 Words
    • 5 Pages
    Great Essays
  • Improved Essays

    Suhani Shah & Emma Bekhet Purpose: To develop an efficient and accurate method of using torque to mass a meter stick – without using a triple beam balance or spring scale / force meter. Materials: One meter stick One fulcrum Three sliding Clips One 50 gram hanging mass Procedures: Setup fulcrum Attach a sliding clip at the 50 cm mark of the meter stick and place meter stick on a fulcrum. From that point, move the sliding clip left and right (based on which side of the meter stick is going down/up) until you are able to identify a precise value for the center of mass.…

    • 1296 Words
    • 6 Pages
    Improved Essays