Which x ray unit has a maximum output of 150 kV?
|
Magnaflux (A/E 99D-1) 150kVP
|
|
what are the power requirements for the magnaflux?
|
115 +/- 10 VAC 50/60 Hz single phase
|
|
What are the major components of the magnaflux?
|
1) master control panel
2) tubehead 3) cooler unit |
|
What is the kV range and working range of the magnaflux
|
0 to 150kV in kV increments and kV starts working at 20 to 150 kV
|
|
What is the MA range of the Magnaflux?
|
1 to 7 MA in .1 increments
|
|
What size is the focal spot on the magnaflux and Lorad?
|
1.5 mm
|
|
What is the continuous operating time for the magnaflux?
|
0 to 99.9 min(99min and 54 secs)
|
|
What controls the quantity of xrays on the magnaflux?
|
the MA control
|
|
What controls the quality and intensity on the magnaflux?
|
the kV control
|
|
What temperature will the thermal cutout switch disable the system?
|
when the anode exceeds 200 degrees
|
|
What psi of SF6 gas is charged into the tube head for the magnaflux?
|
50 PSI
|
|
What is the max PSI set at for the pressure relief valve on the magnaflux?
|
70 PSI
|
|
The pressure switch on the magnaflux will disable the system if the PSI falls below what?
|
20 PSI
|
|
What is the fluid capacity of the cooler unit for the magnaflux and how much is stored in the well
|
5qt capacity with 2qts in the well
|
|
What is the operating pressure of the cooler unit for the magnaflux?
|
50 to 80 PSI
|
|
What are the major components of the Lorad 160 kVP x ray unit?
|
master control panel and tubehead
|
|
What are the power requirements for the lorad?
|
115 +/- 10 or 225+/-25 VAC 50/60 Hz single phase
|
|
What is the kV range of the Lorad?
|
0 to 160kV in 1 kV increments and starts working at 5 to 160kV
|
|
What is the MA range of the Lorad?
|
.0 to 5.0 in .1 MA increments
|
|
What is the continuous operating time for the lorad?
|
0 to 99 mins and 59 secs
|
|
What is the the degrees cone of emission for the magnaflux and lorad
|
40 degrees
|
|
How is the tubehead cooled on the lorad?
|
air cooled by fan
|
|
What is the radiation output for the lorad?
|
240 R/minute at 18 inches(50cm) from target
|
|
What is the normal operating SF6 gas pressure for the lorad?
|
50 psi at 70 degrees
|
|
The lorad's pressure sensor will shutdown the unit if it falls below what?
|
25 PSI
|
|
The lorad's anode thermal sensor will shut down the unit if the anode temp rises above what?
|
220 degrees F
|
|
At what pressure will the lorad's pressure relief valve release SF6 gas from the tube head?
|
between 75 to 80 psi
|