• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/50

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

50 Cards in this Set

  • Front
  • Back
1. What is Pacing that is not inhibited by intrinsic beats
Asynchronous
2. When a Pacing stimulus is delivered between the tip and the ring electrodes on a pacing lead it is known as _____ pacing?
Bipolar pacing
3. Sensing of intrinsic electrical activity between the tip and the
ring electrodes of a pacing lead is known as _______ sensing?
Bipolar Sensing
4. The actual excitation or depolarization of cardiac tissue by a pacing
stimulus is known as?
Capture
5. Reasons why a pacing stimulus may "fail to capture"?
Pacing stimulus is too weak
The tissue is refractory
6. The lead input and circuitry on a pacemaker dedicated to a particular
chamber of the heart.
Channel
7.What is the initial state of activation of cardiac tissue?
Depolarized
8. Cardiac tissue is said to be ____ when it can be activated by a
pacing stimulus or an intrinsic wave of depolarization.
Excitable
9. The opposite of "Excitable" is?
Refractory
10. What is A beat or series of beats that result only from pacing? (in contrast intrinsically conducted beats)
Fully Paced Beats
11. Beats that result from a combination of a paced stimulus and intrinsic activation occur simultaneously are known as?
fused beats
12.What is the absence of intrinsic conduction between the atria and the
ventricles?
Heart block
14. P wave or QRS complex arising from the heart’s own electrical activity, in contrast to a paced beat is known as?
Intrinsic beat (also intrinsic depolarization, intrinsic complex)
15. The patients own Conduction of an atrial impulse (sensed or paced) via the His-Purkinje system, causing a ventricular QRS complex.
Intrinsic conduction
16. The insulated wire that runs from the pacemaker or defibrillator pulse
generator into the heart.
Lead
17. A description of the general way a pacemaker is programmed to pace.
This is described by a code of letters, which specifies which chambers are
paced, and in what ways they are paced.
Mode
18. Sensing on a channel of a pacemaker of something other than
what that channel is supposed to sense.
Oversensing
19. Examples of oversensing by a ventricular channel?
The pacemaker interpreted one of these
signals as QRS complexes causing ventricular oversensing:

senses the T wave
senses an atrial pacing stimulus
senses electrical noise from a lead fracture
20. Atrial activity, on the surface ECG, or as sensed by the pacemaker or defibrillator.
P wave
21.The delay between the delivery of the atrial pacing stimulus and the delivery of the ventricular pacing stimulus.
Paced AV delay
22. A P wave or QRS complex initiated by pacing, in contrast to an intrinsic beat.
Paced beat
23. The period after a ventricular
sensed or paced event, when the atrial channel is refractory.
(PVARP) Post ventricular atrial refractory period
24. The state of the cardiac cell membrane prior to activation
(depolarization). This refers to the separation of positive and negative chargeson the outside and inside of the cell, respectively.
Polarized
25. A pacemaker parameter that can be altered by using the manufacturer’s programmer is said to be?
Programmable
26. A portable device that allows communication with the pacemaker
or defibrillator via an electromagnetic telemetry link. This link allows
downloading of stored information from the pacemaker or defibrillator,and programming of various parameters
Programmer
27. The main body of the pacemaker or defibrillator that houses the battery and circuits.
Pulse Generator
28.The intrinsic QRS complex, as sensed by the pacemaker or defibrillator.
The R wave
29. Cardiac tissue is said to be ______ when it cannot be activated by
a pacing stimulus or an intrinsic wave of depolarization. The opposite of this
is excitable.
Refractory
30. The period of time after a depolarization when the tissue cannot be electrically excited. When the term is applied to a pacemaker, it refers to the period after a sensed or paced beat during which the pacemaker will ignore another sensed event. This is a programmable parameter.
Refractory Period
31. Electrical activity movingin the direction opposite to normal
(which is anterograde).
Retrograde
32. An intrinsic R wave or P wave that has been registered by the pacemaker
or defibrillator. Similar to intrinsic beat, but with the added requirement
that the pacemaker has actually sensed (or registered) the beat.
Sensed beat
33. The function of detecting the intrinsic electrical activity of the heart.
Sensing
34. A description of the level of the sensing function. The setting of
this function gives the minimum amplitude of intrinsic electrical activity that the pacemaker will register.
Sensitivity
35. True or False
If the sensitivity in the ventricular chamber of the pacemaker is set to 5 mV, the intrinsic R wave must transmit an electrical signal at least 5 mV in amplitude for the pacemaker to register that an intrinsic beat has occurred.
True
36. The way a pacemaker is programmed.
Settings
37.The electrical output delivered by a pacemaker, through a pacemaker
lead to the heart, in order to pace the heart.
Stimulus
38. The mark created by a pacemaker stimulus on the surface ECG or telemetry monitor.
Stimulus Artifact
39. A normal behavior of dual chamber pacing in a triggered mode. The
pacemaker senses activity in one chamber (usually the atrium) and delivers a pacing stimulus in the other chamber (usually the ventricle) after a certain time delay (the AV delay).
Tracking
40. A pacing mode in which a sensed beat triggers a paced beat.
This is most commonly used in a dual chamber pacemaker, so that a sensed
atrial beat triggers a paced ventricular beat, after an adjustable delay. In this
setting, the ventricle is said to be tracking the atrium.
Triggered Mode
41.Pacing stimulus delivery between the electrode of pacing lead and the shell of the pulse generator.
Unipolar pacing
42. Sensing of intrinsic electrical activity between the electrode of a pacing lead and the shell of the pulse generator.
Unipolar sensing
43. A short interval of time during which the pacemaker cannot sense any events. The first part of most refractory periods in demand pacemakers is a ________.
Blanking period
44. The active propagation of a depolarization wave in the heart.
Conduction
45. Heart cells pass on an electrical wave from one cell to the next. Some conduct the message faster than others, but virtually all heart cells have this property.
Conduction
46. A phenomenon that can occur in dual chamber pacemakers in which a stimulus or intrinsic event from one chamber is sensed by the other chamber (e.g., an atrial pacing output being sensed by the ventricular lead, or vice versa), resulting in an inappropriate pacemaker response such as inhibition or resetting of the refractory period
Cross talk
47.The area of complex
interaction between the pacing lead’s activeelectrode and the heart tissue.
Electrode Tissue interface
48. The total opposition to
electrical current flow in the pacing circuit measured in ohms (Ω).
Lead impedence
49. The duration of the pacing impulse, expressed in milliseconds.
Pulse width (pulse duration)
50. The negative pole in an electrical circuit. Usually the lead electrode (tip) in a pacing system.
Cathode
51. The positive electrode of an electrical circuit.
Anode