What test tool(s) are used to evaluate Effective Focal Spot Size?
|
Line pair resolution tool
star test pattern pinhole camera >0.3mm |
|
What are the acceptable parameters for effective focal spot size?
|
FSS<0.8mm +50%
0.8</=FSS</=1.5 +40% FSS>1.5mm +30% |
|
How often should FSS be evaluated?
|
on installation of a new x-ray tube and annually thereafter
|
|
What test tool(s) are used to evaluate Half Value Layer?
|
Al Filtration
Digital Dosimeter |
|
What are the acceptable parameters for Half Value Layer?
|
@80 KVp
2.3</= HVL</=3.5 |
|
How often should HVL be evaluated?
|
Semi Annually
|
|
What test tool(s) are used to evaluate Collimation?
|
Collimator test tool (10X12 cassette modified with radiopaque markers)
Beam alignment test tool |
|
What are the acceptable parameters for collimation?
|
Light/rad field alignment:
2% FFD Centering: 1% FFD |
|
How often should collimation be evaluated?
|
semi annually
|
|
What test tool(s) are used to evaluate the Collimator indicator?
|
Ruler, cassette or fluorescent screen
|
|
What are the acceptable parameters for the collimator indicator?
|
+/= 2% of collimator indicator
|
|
How often should the collimator indicator be evaluated?
|
semi annually
|
|
What test tool(s) are used to evaluate distance?
|
tape measure
|
|
What are the acceptable parameters for the collimator indicator?
|
+/= 10% FFD
|
|
How often should distance be evaluated?
|
semi annually
|
|
What test tool(s) are used to evaluate the centering indicator?
|
centering tool
|
|
What are the acceptable parameters for the centering indicator?
|
+/= 2% FFD
|
|
How often should the centering indicator be evaluated?
|
semi annually
|
|
What test tool(s) are used to evaluate angulation?
|
protractor
|
|
What are the acceptable parameters for angulation?
|
+/= 1% FFD
|
|
How often should the angulation be evaluated?
|
semi annually
|
|
What test tool(s) are used to evaluate KVp?
|
Wisconsin (KVp) test cassette
mammo-KVp meter |
|
What are the acceptable parameters for KVp?
|
+/= 5 KVp
Mammo - +/= 1 KVp |
|
How often should KVp be evaluated?
|
semi annually
|
|
What test tool(s) are used to evaluate the timer?
|
Synchronous spinning top
computerized dosimeter |
|
What are the acceptable parameters for the timer?
|
+/= 5% > 10 ms
[+/=20% < 10 ms, Bushong] |
|
How often should the timer be evaluated?
|
at least annually
|
|
What test tool(s) are used to evaluate the mA linearity [mR/mAs]?
|
computerized dosimeter
|
|
What are the acceptable parameters for mA linearity?
|
+/= 10% of adjacent mA stations
|
|
How often should mA linearity be evaluated?
|
at least annually
|
|
What test tool(s) are used to evaluate exposure reproducibility?
|
computerized dosimeter
|
|
What are the acceptable parameters for exposure reproducibility?
|
+/= 5%
|
|
How often should exposure reproducibility be evaluated?
|
at least annually
|
|
What test tool(s) are used to evaluate exposure reproducibility for exposure control?
|
step wedge and dosimeter
|
|
What are the acceptable parameters for exposure control reproducibility?
|
+/= 0.1 OD
|
|
How often should exposure control reproducibility be evaluated?
|
semi annually
|
|
What test tool(s) are used to evaluate exposure control ion chamber sensitivity?
|
Step wedge and dosimeter
|
|
What are the acceptable parameters for exposure control ion chamber sensitivity?
|
+/= 10%
|
|
How often should exposure control ion chamber sensitivity be evaluated?
|
semi annually
|
|
What test tool(s) are used to evaluate exposure control density control?
|
computerized dosimeter
|
|
What are the acceptable parameters for exposure control density control?
|
verification to manufacturer
|
|
How often should exposure control ion chamber sensitivity be evaluated?
|
semi annually
|
|
What test tool(s) are used to evaluate exposure control min response time?
|
Lucite phantom and computerized dosimeter
|
|
What are the acceptable parameters for exposure control min response time?
|
+/= 10% of one another
|
|
How often should exposure control min response time be evaluated?
|
semi annually
|
|
What test tool(s) are used to evaluate exposure control back up timer
|
computerized dosimeter
|
|
What are the acceptable parameters for exposure control min response time?
|
</=600 mAs or 6s (>50KVp)
</=2000 mAs (<50KVp) |
|
What are the parameters for exposure rate on fluoroscopic unit?
|
should be <5 rad/min, but must <10 rad/min (with ABS)
|
|
What are the minimum Source-to-skin distances for mobile and fixed fluoroscopic units?
|
>/= 15" fixed
>/= 12" mobile |
|
What test tool(s) are used to evaluate fluoroscopic resolution for unit and monitor?
|
fluoro wire mesh test tool
|
|
What test tool(s) are used to evaluate fluoroscopic contrast?
|
low resolution test tool
|
|
If the HVL increases, what is it an indication of?
|
The deposition of vaporized tungsten on the inner surface of the glass envelope, or an increase in KV
|
|
If HVL decreases, what is it an indication of?
|
A decrease in the actual KV
|