contact therapy units

40-50 kv


used for surface lesions, skin cancer

Superficial Therapy units

Range of 50 - 150 kv


treatment surface in contact with cone


used aluminum filters to increase penetration



Orthovoltage machines

Range of 150 - 500 kv


deep therapy machines


replaced by linear accelerators

Megavoltage machines

Operate at 1000 kv or more


Van de Graaff, betatron, and linear accelerators


High dose rates


compact design


high reliability

Define Penumbra

It is the beam edge unsharpness, exemplied as scatter

How is a Bending Magnet used?

To change the direction of the accelerated electron beam from horizontal to vertical. Bends the beam

xray target

Accelerator target that generated x-rays are transmitted though to reach the patient. e.g. tungsten.

vacuum pump

to prevent power loss and electrical arcing caused by interactions of electrons with gas molecules



flattening filter (proton beam)

used to create a beam of sufficient area and uniformity

scattering foil (electron beam)

used to provide a uniform beam of dimensions suitable for treatment

Monitor Ionization Chamber

where an ionization current is produced and conducted to the control panel and displayed as monitor units

parallel opposed fields

180 degree images (ex ap/pa, right lateral, left lateral

Bolus

Acts as an extra layer of skin to bring the dose closer to the surface. ex. wet gauze, wet paper towels. It is not skin sparing..



3 things you cannot treat without

Pathology report


Informed Consent


Signed Prescription



Record & Verify

Safeguard system to check parameters to ensure they match the planned treatment

Fractionation

is the individual treatment dose



Protraction

is the time period in which the dose is delivered

When does QA take place?

every 5th treatment



what is the Isocenter?

The point where all three points meet, most likely inside the body

SSD (Source to Skin Distance)

The distance from the source to the patients skin, it is measured by the ODI (Optical Distance Indicator. Most machines are 100cm SAD (Source to Axis Distance)

Multi-leaf Collimators (MLCs)

able to create field shapes w/o the use of focus blocks

What are focus blocks made out of?

BLT with Cheese (bismuth, lead, tin, and cadmium)

Difference btwn electron and photon treatments?

Electron treatments only penetrate a certain depth and die off, photon treatments penetrate all the way through.