20 Cards in this Set

  • Front
  • Back
tech does not have to pick length of exposure therefore does not pick mAs
AEC, phototiming
length of exposure means what
AEC is considered a timing device (can pick mA, doing so influences time)
if low mA is picked exposure will be
long
if high mA is picked exposure will be
short
if AEC is used techs can pick all what other technical factors
kV, focal spot, distance, field size
what are the most common errors when using AEC
incorrect positioning with regard to select chamber/cell, incorrect cell positioning
are the cells ionized chambers
yes
does radiation to the chambers create a charge
yes
t or f, depending on the body part selected, when the chamber reaches a predetermined charge, it will break the timing circuit and stop exposure
true
how are AEC systems calibrated
phantoms and then clinical trials
when multiple cells are used, the sum of all the cells selected is ____ by the ____ of cells in use. termination occurs when the desired charged is ____
divided, number, reached
where are the chambers
between the patient and ir, on the upright Bucky and table buckey
how many chambers chosen for KUB how many for PA chest
3, 2
what settings will adjust the amount of density seen on the image
density controls, -2, -1,N, +1,+2
chambers are calibrated to achieve ____ density in normal settings
optimal
do you change density controls for different body habiti, ex. asthenic to hypersthenic
no
timer set to avoid extreme over exposure in case of some sort of error, such as cell selection, positioning, ext
back up timer
if collimator is sitting over a selected cell, what will happen due to AEC trying to penetrate a dense body part
over exposure
excessive lack of collimation produced excess
scatter
if to many scatter photons reach the selected chamber the exposure will
terminate, EI number will indicate over exposure