• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/78

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

78 Cards in this Set

  • Front
  • Back
The functional images differ from standard nuclear medicine is that they show
metabolism and blood flow on a cellular level.
PET produces
cross sectional functional images of the body.
fMRI/functional MRI
Is the other only modality which also shows function but this is mostly based on regional changes in blood flow.
FDG
Fluorodeoxyglucose
Finger Exposures
Tc-99m
I-131
F-18
4mrem/sec
3mrem/sec
19mrem/sec
What is PET?CT used for?
oncology
cardiology
neurology
What are 2 ways proton rich isotopes turn form unstable to stable....protons to neutrons???
1. Positron emission
2. Electron Capture

Isotopes with too many protons are unstable
Interaction with an electron causes
annihilation
PET scanners are only looking for
511 keV photons.

1.02 MeV or 2 photons at 511 keV.
The fundamental basis for a PET scanning is
annihilation coincidence detection.
ACD
annihilation coincidence detection
Why is collimation in PET not required?
Due to ACD or annihilation coincidence detection
In PET scanners, groups of
detectors are arranged in a ring around the patient. A group of crystals will be viewed by 1 PMT.
Septal rings
are lead shields placed between the detectors to eliminate cross talk between detectors.
When the septal rings are removed......
scatter increases.
LOR/Line of Responses
When 2 events are detected in coincidence, a straight line is drawn between the 2 detectors registering the event.
Why is a high count rate required for PET?
For every coincidence event detected, at least 100 unmatched events are discarded.
In order for the PC to record an event......
1. Two events simultaneously
2. Energy information is recorded...position in pulse height analyzer
Detected events =
T + R + S
True
good data
Random
occur when two unrelated events are detected within the coincidence window. Randoms are estimated and corrected for
Scatter
is usually of lower energy below 511 KeV and can be removed with absorbers placed in front of the crystals.
As we increase the activity......
eventually the randoms will exceed the number of true coincidence events.
Coincidence Detection
The rate of random event detection is also greatly effected by the width of the time window. A larger time window allows more randoms to be detected.
2D
LOR/ lines of response are oimited to the same detector ring or adjacent detectors. Thus, shielding between rings is used.
2D limits counting and
lowers sensitivity.
3D
any LOR is permitted and NO SHIELDING BETWEEN RINGS IS USED.
3D allows for 5 times more sensitivity however,
randoms and scatter increase as well.....
New detectors had to be made due to
the 511 keV penetrating photons. Improved stopping power and detector efficiency.
Heavier elements and are better photon attenuators
BGO
LSO
GSO
Energy window and detector size
effect detector sensitivity.
BGO has the greatest sensitivity.
Since no collimation is used in PET.......
sensitivity is 10 to 20 times greater and allows for a lower dose adm. patient.
Scatter degrades the
image
Compton Scatter
for lower atomic numbers materials
Photoelectric absorption
for higher atomic numbers. Thus, BGO crystals with high density and high atomic number reduces the number of scatter events in the crystals
Inside a person
Compton scatter
Inside a BGO crystal
photoelectric effect
Light yield
the amount of light produced per unit energy. NaI(Tl) crystals have high light yield despite their low stopping power
The light produced by different detectors crystals varies in
amount, wavelength and time of decay
Time of decay is
the amount of time it takes for light output to reach zero after a scintillation is initiated.
A _______ decay time desirable so the detector can produce more scintillations in a short time
fast
time of decay is related to the ________ ______ of the detector,
dead time
BGO has a _______
long decay time of 300 nsec
Spatial Resolution on A PET
2 limits exist due to....
1. range of positrons in tissues
2. The fact that the 180 degrees separation of positron annihilation photons is not exactly true.
1). Positrons can travel in tissue just as beta particles do....their range if a function to their energy so
with higher energy positrons travel further
For Fluorine 18, positrons have a maximum range of 2.6 mm however the actual effect on resolution is less than
1 mm
2.) Annihilation photons deviate from 180 degrees path because at the moment of annihilation,
the positron-electron pair is not completely at rest. The result is an angular spead.
Therefore, the combined action of these two effects limit PET spatial resolution to
1.5-3.0 mm
These effects and detector resolution together determine the system resolution. Modern PET scanners are capable of final image resolution in the range of
8 mm
Attenuation Correction
In PET scanning, a transmission scan is performed to correct for the patient attenuation of different densities in the patient.
This is an effective method since 5 11 photons travel in a
straight line.
By measuring the total attenuation along the line of travel
attenuation can be corrected from the emission image.
Dedicated PET scanners use internal sources how ever, in a PET/CT
the CT scan is used for attenuation correction.
Daily QA PET
A daily ________ scan is required.
blank = This is similar to a flood on the camera. A uniform source is used to expose the PET detectors. This may be internal rods sources or a cylinder.
The purpose of the blank scan is to determine if the scanner is OK for clinical use.
More specifically, the scan can be used to monitor system stability and to determine which detectors are more or less sensitive then average
Normal uptake of FDG
Brain, Breast, Bowel, Heart, Liver, Salivary glands, stomach, thyroid, urinary
Brain
high uptake in gray matter due of metabolic substrate.
Myocardial uptake
variable, depends on substrate availability
Kidney
route of excretion (filtered, poorly absorbed from renal tubules and excreted unchanged.)
Muscles
variable depends on muscle tension or activity shortly or during FDG uptake phase
Bone marrow
mild upatke
liver
high initial uptake with rapid clearance due to relatively high glucose -6-phosphatase activity
lungs
low uptake (low initial extraction of FDG)
GI tract
variable uptake
Mouth, oropharynx,pharynx
mild to moderate uptake
Breast, thyroid, thymus,
mild to moderate uptake
BIOdistribution of Fdg
brain = 6.9 % OF INJ. DOSE
liver = 4.4
heart = 3.3
red marrow = 1,7
kidneys = 1.3
lungs = 0.9
Breast Uptake
- Breast blushing
especially in patient who breast-feed
resume breast feeding after 10 t/12
Recent Surgery
will show uptake
Non-fasting patient
will have bowel uptake
hip replacement and patient exercise
will show uptake in muscle activity
PET Artifacts
Brown fat can take up FDG. Shivering can cause uptake in brown fat so it is important to keep the patient warm during the uptake period.
Muscles will also take up FGD and patient should remain calm and still during the uptake period.
Strenuous exercise should be avoided 24 hrs prior to exam. Shivering and talking may also cause muscular activity.
Smokers may exhibit uptake in
the intercostal muscles
Patient chewing gum will
show uptake.
Artifacts can also be caused by mismatch in the PET vs CT scan due to motion, esp. .....
respiration scans of the thorax
- This can cause a curvilinear cold band parallel to the dome of the diaphram.
High density objects in the CT will cause over correction in the PET scan making false hot spots
Common causes are dental implants, cardiac pacemakers, central intravenous lines, and metallic limb protheses.
Oral or intravenous contrast can cause
errors in attenuation correction.