Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
78 Cards in this Set
- Front
- Back
The functional images differ from standard nuclear medicine is that they show
|
metabolism and blood flow on a cellular level.
|
|
PET produces
|
cross sectional functional images of the body.
|
|
fMRI/functional MRI
|
Is the other only modality which also shows function but this is mostly based on regional changes in blood flow.
|
|
FDG
|
Fluorodeoxyglucose
|
|
Finger Exposures
Tc-99m I-131 F-18 |
4mrem/sec
3mrem/sec 19mrem/sec |
|
What is PET?CT used for?
|
oncology
cardiology neurology |
|
What are 2 ways proton rich isotopes turn form unstable to stable....protons to neutrons???
|
1. Positron emission
2. Electron Capture Isotopes with too many protons are unstable |
|
Interaction with an electron causes
|
annihilation
|
|
PET scanners are only looking for
|
511 keV photons.
1.02 MeV or 2 photons at 511 keV. |
|
The fundamental basis for a PET scanning is
|
annihilation coincidence detection.
|
|
ACD
|
annihilation coincidence detection
|
|
Why is collimation in PET not required?
|
Due to ACD or annihilation coincidence detection
|
|
In PET scanners, groups of
|
detectors are arranged in a ring around the patient. A group of crystals will be viewed by 1 PMT.
|
|
Septal rings
|
are lead shields placed between the detectors to eliminate cross talk between detectors.
|
|
When the septal rings are removed......
|
scatter increases.
|
|
LOR/Line of Responses
|
When 2 events are detected in coincidence, a straight line is drawn between the 2 detectors registering the event.
|
|
Why is a high count rate required for PET?
|
For every coincidence event detected, at least 100 unmatched events are discarded.
|
|
In order for the PC to record an event......
|
1. Two events simultaneously
2. Energy information is recorded...position in pulse height analyzer |
|
Detected events =
|
T + R + S
|
|
True
|
good data
|
|
Random
|
occur when two unrelated events are detected within the coincidence window. Randoms are estimated and corrected for
|
|
Scatter
|
is usually of lower energy below 511 KeV and can be removed with absorbers placed in front of the crystals.
|
|
As we increase the activity......
|
eventually the randoms will exceed the number of true coincidence events.
|
|
Coincidence Detection
|
The rate of random event detection is also greatly effected by the width of the time window. A larger time window allows more randoms to be detected.
|
|
2D
|
LOR/ lines of response are oimited to the same detector ring or adjacent detectors. Thus, shielding between rings is used.
|
|
2D limits counting and
|
lowers sensitivity.
|
|
3D
|
any LOR is permitted and NO SHIELDING BETWEEN RINGS IS USED.
|
|
3D allows for 5 times more sensitivity however,
|
randoms and scatter increase as well.....
|
|
New detectors had to be made due to
|
the 511 keV penetrating photons. Improved stopping power and detector efficiency.
|
|
Heavier elements and are better photon attenuators
|
BGO
LSO GSO |
|
Energy window and detector size
|
effect detector sensitivity.
BGO has the greatest sensitivity. |
|
Since no collimation is used in PET.......
|
sensitivity is 10 to 20 times greater and allows for a lower dose adm. patient.
|
|
Scatter degrades the
|
image
|
|
Compton Scatter
|
for lower atomic numbers materials
|
|
Photoelectric absorption
|
for higher atomic numbers. Thus, BGO crystals with high density and high atomic number reduces the number of scatter events in the crystals
|
|
Inside a person
|
Compton scatter
|
|
Inside a BGO crystal
|
photoelectric effect
|
|
Light yield
|
the amount of light produced per unit energy. NaI(Tl) crystals have high light yield despite their low stopping power
|
|
The light produced by different detectors crystals varies in
|
amount, wavelength and time of decay
|
|
Time of decay is
|
the amount of time it takes for light output to reach zero after a scintillation is initiated.
|
|
A _______ decay time desirable so the detector can produce more scintillations in a short time
|
fast
|
|
time of decay is related to the ________ ______ of the detector,
|
dead time
|
|
BGO has a _______
|
long decay time of 300 nsec
|
|
Spatial Resolution on A PET
2 limits exist due to.... |
1. range of positrons in tissues
2. The fact that the 180 degrees separation of positron annihilation photons is not exactly true. |
|
1). Positrons can travel in tissue just as beta particles do....their range if a function to their energy so
|
with higher energy positrons travel further
|
|
For Fluorine 18, positrons have a maximum range of 2.6 mm however the actual effect on resolution is less than
|
1 mm
|
|
2.) Annihilation photons deviate from 180 degrees path because at the moment of annihilation,
|
the positron-electron pair is not completely at rest. The result is an angular spead.
|
|
Therefore, the combined action of these two effects limit PET spatial resolution to
|
1.5-3.0 mm
|
|
These effects and detector resolution together determine the system resolution. Modern PET scanners are capable of final image resolution in the range of
|
8 mm
|
|
Attenuation Correction
|
In PET scanning, a transmission scan is performed to correct for the patient attenuation of different densities in the patient.
|
|
This is an effective method since 5 11 photons travel in a
|
straight line.
|
|
By measuring the total attenuation along the line of travel
|
attenuation can be corrected from the emission image.
|
|
Dedicated PET scanners use internal sources how ever, in a PET/CT
|
the CT scan is used for attenuation correction.
|
|
Daily QA PET
A daily ________ scan is required. |
blank = This is similar to a flood on the camera. A uniform source is used to expose the PET detectors. This may be internal rods sources or a cylinder.
|
|
The purpose of the blank scan is to determine if the scanner is OK for clinical use.
|
More specifically, the scan can be used to monitor system stability and to determine which detectors are more or less sensitive then average
|
|
Normal uptake of FDG
|
Brain, Breast, Bowel, Heart, Liver, Salivary glands, stomach, thyroid, urinary
|
|
Brain
|
high uptake in gray matter due of metabolic substrate.
|
|
Myocardial uptake
|
variable, depends on substrate availability
|
|
Kidney
|
route of excretion (filtered, poorly absorbed from renal tubules and excreted unchanged.)
|
|
Muscles
|
variable depends on muscle tension or activity shortly or during FDG uptake phase
|
|
Bone marrow
|
mild upatke
|
|
liver
|
high initial uptake with rapid clearance due to relatively high glucose -6-phosphatase activity
|
|
lungs
|
low uptake (low initial extraction of FDG)
|
|
GI tract
|
variable uptake
|
|
Mouth, oropharynx,pharynx
|
mild to moderate uptake
|
|
Breast, thyroid, thymus,
|
mild to moderate uptake
|
|
BIOdistribution of Fdg
|
brain = 6.9 % OF INJ. DOSE
liver = 4.4 heart = 3.3 red marrow = 1,7 kidneys = 1.3 lungs = 0.9 |
|
Breast Uptake
|
- Breast blushing
especially in patient who breast-feed resume breast feeding after 10 t/12 |
|
Recent Surgery
|
will show uptake
|
|
Non-fasting patient
|
will have bowel uptake
|
|
hip replacement and patient exercise
|
will show uptake in muscle activity
|
|
PET Artifacts
|
Brown fat can take up FDG. Shivering can cause uptake in brown fat so it is important to keep the patient warm during the uptake period.
|
|
Muscles will also take up FGD and patient should remain calm and still during the uptake period.
|
Strenuous exercise should be avoided 24 hrs prior to exam. Shivering and talking may also cause muscular activity.
|
|
Smokers may exhibit uptake in
|
the intercostal muscles
|
|
Patient chewing gum will
|
show uptake.
|
|
Artifacts can also be caused by mismatch in the PET vs CT scan due to motion, esp. .....
|
respiration scans of the thorax
- This can cause a curvilinear cold band parallel to the dome of the diaphram. |
|
High density objects in the CT will cause over correction in the PET scan making false hot spots
|
Common causes are dental implants, cardiac pacemakers, central intravenous lines, and metallic limb protheses.
|
|
Oral or intravenous contrast can cause
|
errors in attenuation correction.
|