• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/162

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

162 Cards in this Set

  • Front
  • Back
What does the O2 flush valve deliver? and where does it go?
35-75L/min of 100% O2; to the Common Gas Outlet
Whats the purpose of the Fail Safe system? How does it work?
-halts the flow of all gasses except Oxygen in the event of O2 failure.
-valve/"gate" held open by pressure in O2 line. Only allows flow of other gases if O2 pressure is sufficient.
When using Piston Ventilators, will the TV delivered by affected by variation in O2 pipeline pressure?
NO- piston ventilators use electric motors to compress bellows and deliver TV.
Function of the proportioning system is...
Prevent delivery of hypoxic mixture (makes sure ratio of N2O to O2 is not more than 3:1). Links N2O and O2 mechanically or pneumatically, to prevent final O2 concentration less than 25%. If N2O increased beyond max allowed there is a proportional increase in O2. If O2 flow lowered, proportional decrease in N2O.
In what 4 situations could a hypoxic mixture be delivered despite use of hypoxic safeguard?
1) wrong supply gas in tank or pipeline
2)defective pneumatics or mechanics
3)leaks downstream of flow control valve
4)inert gas admin (ex. 3rd gas like helium)
Rate of vaporization depends on what factors?
1) temperature
2) vapor pressure of the liquid
3)partial pressure of the vapor above the evaporating liquid
What is the splitting ratio?
How is it determined?
It's the gas entering the vaporizer chamber divided by the FGF.
-automatically determined in variable bypass vaporizers by internal resistance to flow when a specific % is selected
During Sevoflurane administration an unusually delayed rise or unexpected decline of inspired concentration comparted to vaporizer setting may be associated with what?
Excessive heating of the CO2 absorbant cannister
Resistance in a breathing circuit can be minimized by...
-reducing circuit length
-increasing diameter of tubing
-avoiding use of sharp bends (keep angles less than 20 deg)
-eliminating valves
-maintaining laminar flow
Advantages of rebreathing
-economical
-Increase or maintain tracheal warmth and humidity
-decrease potential for exposure of gasses to OR personnel
-maintain pt temp
-prevent soda lime granules from drying out
Rebreathing of exhaled Nitrogen slows induction true or false?
True
Common features of Non-rebreathing systems include:
-Lack unidirectional valves
-Lack CO2 absorbant
-Amount of rebreathing is HIGHLY dependent on FGF rate
-Resistance is kept to a minimum
Two common reasons for an increase in inspired CO2 with circle system?
1) Exhausted absorbant granules
2) Faulty unidirectional valves
True or False: Incompetence of inspiratory or expiratory valves turns the entire corrugated limb into dead space.
True
How do you convert a closed circle system to a semi open configuration and why would you do this?
-Increase FGF to 5-8L/min and open APL valve
-Do this if inspired CO2 of more than 1-3mmHg is detected on capnograph
True/False: Oxyhemoglobin absorbs a greater amount of Red light.
False: it absorbs a greater amount of infrared light at 960nm.
True/False: Methemoglobin has the same absorption coefficient at both red (660) and infrared light (940).
True
True/false: Carboxyhemoglobin and Oxyhgb absorb light at 660nm identically.
True
Name a situation in which a patients sat of 90% MAY indicate a PaO2 of LESS than 65mmHg.
-hypothermia
-alkalosis
-decreased 2,3 DPG
-hypocarbia
(shift to the Left on Oxyhgb dissociation curve)
Laminar flow occurs at------- Reynolds numbers, where
------forces are dominant.
Laminar flow at LOW reynolds numbers, where VISCOUS forces are dominant.
What kind of flow would you expect to see at HIGH reynolds numbers?
Turbulant flow, dominated by inertial forces
For any sized tube transition from laminar straight to turbulent flow occurs when the Re number reaches what value?
Reynolds number of 2100
Flow through an abrupt orifice (ie. flowmeter) becomes turbulent at Re # of what?
Less than 100
What 3 functions does a hanger yoke serve?
1) orients cylinders
2)provides gas tight seal
3) ensures undirectional flow
Whats the calculation to determine how long an O2 cylinder will last?
remaining contents (L)/gauge pressure (psi)=capacity (L)/service pressure (psi)

As oxygen is expended, the cylinder’s pressure falls in proportion to its content
1000psig = E cylinder that is half full =330L at atmospheric pressure and a temp. of 20°C
3L/min= empty in 110 min.
which absorbs more light:
Oxyhemoglobin or deoxyhemoglobin
deoxyhemoglobin- (dark so absorbs more light)
infrared light wave length
960 nm
red light wave length
660 nm
if more 660 light is absorbed...
more deoxyhemoglobin
if more 960 light is absorbed...
more oxygenated hemoglobin
Will the N2O cylinder pressure change as N2O is used?
NOT until the liquid is gone in which case it is more than 3/4 empty. b/c 745psig is the vapor pressure
5 pathways of O2 in the Anesthesia Gas Machine are?
1) FGF flowmeter.
2) Powers O2 flush valve
3) Activates fail safe
4) activates o2 low pressure alarm
5)Compresses bellows of mechanical ventilator.
What does the ASA standard 1 say?
A qualified anesthetist must be present at all times.
What does ASA standard II say?
oxygenation, ventilation, circulation and temperature will be monitored at all times.
Ventilation is assessed by?
-pt assessment (chest rise, breath sounds etc)
-volume of gas (spirometry)
-alarms
-metabolic end products like CO2
Where will you find the most common disconnect in the breathing circuit?
The Y piece
What does ENDCO2 tell us?
Expired level of CO2
-gold standard for evaluating ETT placement
-tells us if the heart is pumping
-helps to determine the quality of ventilation
Name 3 ways to determine CO2?
1) Mass spectrometry
2) Infrared absorption
3) Colorimetric detection
True or False: Hemodynamic instability is associated with hypothermia.
True
What absorbs more light? oxygenated or deoxygenated blood
And at what wavelength?
Deoxygenated (660 nm)
Oxygenated (940 nm)
what causes the oxygen to be released from hemoglobin
enzyme 2,3-DPG
carboxyhemoglobin toxicity
cherry red appearance
methemoglobinemia
blue appearance
SpO2 reading 85%
true or false: If one extremity is above the other, pulse amplitude and SpO2 will be greater in the upper than dependent hand
true
What happens to normal saturations as altitude above sea level decreases?
It decreases
Oxidation product of hemoglobin that forms a reversible complex with oxygen and impairs unloading of oxygen to tissues
Methemoglobin
Whats the reversal agent for Methemoglobinemia?
methylene blue IV 1mg/kg
Signs and symptoms of Methemoglobinemia?
brownish grayish cyanosis
tachypnea
Metabolic acidosis
headache
irritability
What substances may interfere with pulse oximetry and cause erroneous readings?
Black, blue and green nailpolish
Methylene blue, indigo carmine, indocyanine green dyes
How does colorimetric detection work?
exhilation gas expired contains moisture which interacts with CO2 liberating H+ causing acidification which in turn causes the color change from purple to yellow.
-tells you that CO2 is present
-No data to evaluate ventilation.
name 2 reasons capnography waveform would not return to baseline level of 0 with inspiration?
1) pt is rebreathing CO2
2)an obstruction at small airways just proximal to alveoli.
True or False: A good alveolar plateau greatly increases the chances that the EtCO2 reading is a reliable estimate of the level
True
Alterations in ENDTCo2 can be the result of:
Rebreathing CO2
Hypoventilation
Increased CO2 production (hypermetabolic state, MH)
List the basic components of the anesthesia machine.
A source of compressed gases
Flow meters
Vaporizers
A method to manually and mechanical ventilate
scavenger system
List Functions of the Anesthesia Machine.
-Receives medical gases from a gas supply
-Controls the flow of desired gases reducing their pressure
-Vaporize volatile anesthetics into the final gas mixtures
-Deliver the gases to the breathing circuit
Pipeline gas is delivered at what pressure? and by what system?
45-55 psig (pressure on starting machine)
DISS system (Diameter Index Safety System)
What is the purpose of PISS and what does it stand for?
to prevent misconnection of cylinder to the wrong yoke.
PISS (Pin Index Safety System)
Cylinger pressure regulators convert the pressure from what to what?
From Cylinder pressure of 2200psig to 45 psig
Describe the PISS system configuration for the following gasses.
-O2, N2O, air
-O2= 2,5
-N2O= 3,5
-air= 1, 5
Describe 3 functions of the check valve located in the cylinder yoke.
1.   Minimize trans-filling
2.   Allow change of cylinders during use
3. Minimize leaks to atmosphere if a yoke is empty
Describe the Bourdon gauge?
- It measures cylinder pressure
It is a flexible tube within this gauge straightens when exposed to gas pressure, causing a gear mechanism to move a needle pointer
What is the purpose of a secondary regulator? What does it reduce pressures to in the Anesthesia machine?
To further decrease pressures.
Decreases oxygen to 20 psig and nitrous oxide to 38 psig
Whats the function of Proportioning system and how does it work?
Links nitrous oxide and oxygen flows, mechanically or pneumatically, to prevent final inspired oxygen concentration less than 25%
What is another name for the Oxygen supply failure protection device?
Fail safe
To reduce back flow

What Promotes unidirectional flow and is
found on both wall and cylinder gas flow?
Check valves
What happens when you push the O2 flush valve?
-delivers HIGH flow 35-75L of 100% O2 directly to pt.
-dilutes your anesthetic which could potentially lead to awakening
High pressure system consists of what 4 parts?
-parts which receive gas at cylinder pressure (O2=2200psi)
-Hanger yoke (including filter and unidirectional valve)
-Yoke block
-Cylinder pressure gauge
-Cylinder pressure regulators
Intermediate pressure system consists of what parts?
-Receives gases at low constant pressures (37-55 psig, which is pipeline pressure) (Receive gases from regulators or pipeline and delivers to flow controls or oxygen flush valve)
-Pipeline inlets and pressure gauges
-Ventilator power inlet
-Oxygen pressure-failure device (fail-safe) and alarm
-Flow meter valves
-Oxygen second-stage regulator
-Oxygen flush valve
Low Pressure circuit consists of what components?
-includes components distal to the flow meter needle
-Valves
-Flow meter tubes
-Vaporizer
-Check valves (if present)
-Common gas outlet
(Takes gas from flow control valves to common gas outlet)
name 3 types of vaporizers?
-Measured flow: Copper kettle, vernitrol
-Blender: electronic (Ohmeda Tec6-Desflurane vaporizers, Aladin cassette vaporizers)
-modern conventional vaporizer (eg. Variable Bypass Vaporizer)
True or False: Only some of the air flow passing through Copper kettle vaporizer comes in contact with the anesthetic agent.
False: ALL the air that passes through this type of vaporizer comes in contact with anesthetic liquid.
Copper Kettle system is based on what?
FLOW of the anesthetic liquid. Adjust flow on dedicated flow meter up or down to alter concentration of agent.
Name some charachteristics of a Modern Conventional vaporizer.
-Agent specific
-Capable of delivering a constant concentration
Regardless of temperature -changes or flow through the vaporizer occur by
turning a single calibrated control knob counterclockwise to desired percentage
What is the minimum flow recommended on most flow meters and what pressur does it deliver it under?
200-300cc/min
at 16 psig (some books say 20 psig)
Name some potential causes of flow meter malfunction.
-dirt,
-vertical misalignment
-sticking of float at the top of a tube
-leaving flow meter on when machine is turned off then next time machine is turned on float excelerates to top damaging flow meter.
Name the parts of a Thorpe tube?
ie. flow meter, constant pressure variable orifice flow meter.
-Parts: needle valve, indicator float, knobs, valve stops
A secondary regulator drops the pressure to a N2O flow meter to what?
- N2O to 38 psig
Describe what will happen if you attempt to manually ventilate a pt with the APL valve open.
The bag will not fill properly preventing you from getting a seal and delivering manual breaths.
True or False: You can cause Barotrauma if you attempt to manually ventilate your patient with a closed APL valve.
True
Name 3 hazards associated with the APL valve.
-Can cause barotrauma
-accidental activation
-sticking in the on position
Name the components of the Circle system.
-Fresh Gas Flow
-I/E unidirectional valve
-I/E tubing
-Y – connector
-Pop off (overflow)
-Reservoir bag
-CO2 absorbent
What will happen if the inspiratory valve is stuck open on the circle system?
Rebreath CO2 b/c it wont close allowing what pt expires to back flow.
ratings
An assessment of a representative's voting record on issues important to an interest group
What is the current minimum flow recommendation when administering Sevofluratne?
not recommended at total fresh gas flows less than 1 L/min for more than 2 MAC-Hours
What does the scavagening system consist of?
Scavenger System consists of: 1) gas collecting assembly,
2) a transfer means,
3) the interface,
4) gas disposal tubing,
5) gas disposal assembly. (some or all components may be combined). ASTM standard fitting size for scavenger hoses 19 mm ( international standard 30mm)
What is the function of the scavenger interface?
It protects the breathing circuit from excess positive or negative pressure. Must contain a positive pressure relief to vent excess gasses in the event of an obstruction.
What are the National Institutes for Safety and Health's recommendations as far as limiting room concentrations of gasses?
Nitrous oxide to 25ppm
Halogenated agents to 2ppm
Halogenated agents with nitrous 0.5ppm
What does Phase III of the Capnography waveform represent?
The alveolar plateau- emptying of distal alveoli.
American National Standards Institute
Organization who set the stds for the Anesthesia machine
Cylinder pressure regulator
Converts high (~2200), variable pressure to a constant pressure of ~45-47psi downstream of the regulator.
3 types of valves and their functions
Free floating: Moves by gas pushing it (DISS/Hangar Yolk System). Prevents leaking of gas.
Ball and Spring: Permits gas flow by adjoining two pieces (male to female) displacing the ball
Diaphragm Valve:
seen on the 1st/2nd stage regulators
They reduce pressure
Check valve position and purpose (Jackie said to know)
To reduce back flow
Promotes unidirectional flow
Found on both wall and cylinder gas flow
Copper Kettle vaporizer (how works, classified)
Classified as:
A measured flow vaporizer or flow meter-controlled vaporizer
The amount of carrier gas bubbled through the volatile anesthetic is controlled by a dedicated flow meter
All of the gas entering the vaporizer passes through the anesthetic liquid and becomes saturated with vapor
Bourdon Gauge purpose and how it works
Cylinder pressure is measured by the Bourdon gauge
It is a flexible tube within this gauge straightens when exposed to gas pressure, causing a gear mechanism to move a needle pointer
If the lipid solubility of an agent is high then the MAC of the agent is: a. High b. Low
b. low (from sweat book 334)
Potency of an inhalational agent is directly related to its
a. MAC.
b. blood solubility.
c. lipid solubility.
d. vapor pressure.
c. lipid solubility (sweat book 334)
MAC is:
a. inversely related to potency.
b. directly related to potency.
c. inversely related to vapor pressure.
d. directly related to blood solubility.
a. inversely related to potency (sweat book 334)
Potency is directly related to lipid solubility. The oil (o1ive):gas partition coefficient provides a quantitative measure of lipid solubility (olive oil resem-
bles membrane lipids)
The minimum alveolar concentration (MAC) is the alveolar concentration of an inhalational agent at which 50%of patients do not move in response to skin incision or similar noxious stimulus;MAC is inversely related to potency.
If an agent is highly blood soluble, then the blood:gas partition coefficient is:
a. low.
b. high.
c. the same as oil:gas partition coefficient.
d. may be high or low depending on the patient.
b. high (sweat book 334)
Blood solubility determines the speed at which the partial pressure of the agent builds up in the blood and subsequently in the brain. It also determines the speed at which an agent is eliminated from the blood and the brain
The blood:gas partition coefficient provides a quantitative measure of blood solubility
Blood solubility determines:
a. speed of uptake of agent by the brain.
b. MAC.
c. potency.
d. the degree of hypotension the agent will produce
a. speed of uptake of agent by the brain (sweat book 334)
Blood solubility determines the speed at which the partial pressure of the agent builds up in the blood and subsequently in the brain. It also determines the speed at which an agent is eliminated from the blood and the brain
The b1ood:gas partition coefficient provides a quantitative measure of blood solubility
Circle system Unidirectional valves other names
aka flutter valves, one way valves, check valves, directional valves, dome valves. Found on Inspiratory and Expiratory flow ports
Most common ventilator problems
Disconnect most common at y-piece
Precordial
etCO2
Occlusion of breathing circuit: may lead to barotrauma
kinked tube
APL tight when masking
Managing a problem:
Attempt to manual ventilate
Ambu on O2 without machine
The medical gas commonly used are:
Oxygen (green cylinder)
Nitrous oxide (Blue)
Air (Yellow)
Nitrogen (black)
Please note; while technically not a gas, vacuum exhaust for anesthetics waste gas disposal and surgical suction are considered important to the medical gas system
What breathing system is associated with the least amount of resistance?
Open (Insufflation)
eg. Old ether screen
Where is the most resistance located in an intubated patient?
connection between ETT and circuit connector/Y-piece
What is the source of the most resistance in an Unintubated patient?
at expiratory valve in semi-closed system
Density of gas determines flow at ______ flow rates
HIGH
Viscosity of gas determines flow at _____ flow rates
LOW
If you increase the length of tubing what happens to the flow?
You DECREASE the flow
(Poiseulilles Law)
delivery of gases from anesthesia machine via a circuit or mask held above patient’s face
Insufflation
Alternate names for the APL valve
-overflow
-exhaust
Name some advantages of the semi-open system.
-simple design,
- ability to change anesthetic depth rapidly
-portability
- lack of rebreathing (provided fresh gas flow is adequate)
- low resistance
Identify disadvantages of the Semi-open system.
-loss of heat and humidity,
-limited ability to scavenge gases (pollution),
- high gas flow requirements 
-less economical
What are the two main features that CRNA's need to know regarding medical gas systems in order to prevent pt endangerment
Sources of the gases & means of delivery are necessary to prevent and detect medical gas depletion or supply line misconnection
Department of Transportation (DOT) role in Anesthesia
responsible for the regulation of contents of compressed gas cylinders. Must contain: Serial #, service pressure, DOT specification number
What markings are required by DOT?
Serial #, service pressure, DOT specification number
Standard Organizations in Anesthesia
Department of transportation (DOT)
Food & Drug Administration (FDA) National Fire Prevention Association (NFPA)
Compressed Gas Association (CGA)
JCAHO
American Society for testing and Materials (ASTM)
JCAHO role in Anesthesia
Requires preop assessment in all pts - never to OR with no preop assessment
Characteristics of Medical GAS
O2 - 625-700 L capacity, 1800-2200 psi Pressure, USA color-green, international color-white, form-gas
Air - 625-700 L capacity - 1800-2200 psi Pressure, color USA-yellow, color international-white&black, form-gas
N2O-1590 L capacity, 745 psi pressure, USA&intern'l color-blue, form-liquid
oxygen how manufactured, stored
Manufactured by fractional distillation of liquefied air
Stored as a compressed gas at room temperature or refrigerated as a liquid
Stored in banks of H – cylinders connected to a manifold
Manifolds contains valves that reduce cylinder pressure to line pressure
Automatically switch banks when one group is exhausted
Liquid: Must be stored below it’s critical temperature -119°C
The anesthesia provider must always have an emergency (E-cylinder) supply of oxygen available on the anesthesia machine
Nitrous Oxide Characteristics, manufacture, storage, critical temp
Most commonly used anesthetic gas (analgesia & sleepy)
Fills spaces in body (i.e. N/V-prevent via antiemetics)
Manufactured by heating ammonium nitrate
Stored in large high pressure cylinders
Critical temperature 36.5°C
Kept liquid without refrigerator system
If rises above critical temperature it will revert to gaseous phase

Transformation into gaseous phase is not accompanied by great rise in pressure
Relief Valve: other name and what it's designed to do
(aka rupture disk)
Prevent explosion of unexpected high gas pressure
(Ie, designed to rupture at 3300 psig)
Volume of the tank is not proportional to the pressure
Only reliable way to determine residual is to weigh the cylinder
Higher reading indicates:
Gauge malfunction, overfill or cylinder containing a gas other than NO
Air in Anesthesia (reasons for use, critical temp)
Becoming more frequent in anesthesia: keeps integrity of Alveoli d/t nitrogen in air while 100% O2 leads to atelectasis. Also, air used to prevent fire in trach insertion with bovie use.
Critical temperature is – 140.6°C
Pressures fall in proportion to their content
If you tighten your APL valve what do you do to pressure and resistance?
Increases resistance
Normal adult minute volume
5-8L/min
VA (Alveolar ventilation)+ DV(dead space ventilation)=
Minute volume or Minute ventilation
True or False: The higher the minute volume the less carbon dioxide (CO2) the person is releasing.
FALSE: higher the minute volume the more CO2 the patient is releasing.
Do soda lime granules create resistance in the breathing circuit?
Yes. All absorbants add some resistance to the breathing circuit.
Vacuum in OR
Consist of two independent suction pumps
Traps located at every user site which helps to prevent contamination of the system
Effects of environmental temp in OR
Cold pts slow to wake up, wound infection, blood loss, coagulation issues.
Environmental factors of OR
Temperature:
Hypothermia: wound infection, blood loss, coagulation issues
Neurological protection (Crani or CPB); 24°C burn patients
Humidity
Static
Flammable vs nonflammable
Ventilation
Air flow: decreases contamination of the surgical site
Noise
Detrimental on human cognitive functions
Pressures at cylinder, pipeline and fail safe system
2200 psi; 45-55 psi; O2 drops below 25 psi
Max leak current in OR
10 micro Amps
Line isolation Monitor:
determines the degree of isolation between two power wires and the ground and predicts the amount of current that could flow if a short circuit develops. Alarm activates if too high current. Power not interrupted. Upon activation of alarm, last electrical equipment plugged in should be unplugged.
When is there a risk of electrocution in regard to a circuit
Two conductive materials at different voltage potentials may complete a circuit
Protection from electric shock in OR
Isolated transformer: OR power supply isolated from ground
Line isolation monitor: determines the degree of isolation between two power wires and the ground and predicts the amount of current that could flow if a short circuit develops
Another name for a semi-open breathing system.
Flow controlled breathing system.
In a Semi-open breathing system when will rebreathing occur?
When the inspiratory flow exceeds the fresh gas flow.
FLOW is key
Advantages of the Mapleson system.
-Minimal dead space
-Low resistance (great for fragile airway/peds)
-Can be used w/mask and ETT
-Inexpensive
Which Mapleson is Characterized by placement of APL valve near the patient and separation of gas reservoir bag from valve by corrugated tubing?
mapleson A (Magill)
Which Mapleson is most efficient for CO2 elimination for a spontaneously breathing patient?
Mapleson A
Surgical Diathermy: Electrosurgical Units (ESU) names and interaction
Bovie pad: return electrode or exit pathway for electrical current generated through the cautery tip - grounds the pt to prevent spark. placed on muscle. Completes the circuit to the machine
Pacemaker - place bovie pad away from pacemakers
Malfunction of ESU
Bipolar - bovie that has ground contained in tip
Fires & explosions prevention in the OR
~100 fires/yr, most involve face (i.e. chloraprep not dried), suction could be used in the folds of the drape to prevent collecting of gases. Burn injuries include warmed IV bags (burning of vessel), warming blankets.
Vapor Pressure (VP)
Created by molecules in the vapor phase bombarding the walls of a container
Saturated Vapor Pressure (SVP)
Gas phase above the liquid usually at 20°C
↑Temp → ↑SVP
VP is independent of atmospheric pressure if the temperature remains constant
Boiling point
The temperature at which the vapor pressure equals atmospheric pressure
Units of Vapor Concentration
Absolute pressure (mmHg)
Volumes percent
Dalton's law of Partial Pressures
The total pressure exerted by a gas mixture is the sum of the individual pressures of its constituents
Latent Heat of Vaporization
The # of cal required to change 1 g of liquid into vapor without a temp change
Specific Heat
The # of cal required to increase the temp of 1 g of a substance by 1°C
Isoflurane
MAC 1.15%
Sevoflurane
MAC 1.7 %
Desflurane (MAC, vaporizer needs)
MAC 6-7.25; Requires an external heat source
4-9 times more Des is required than other IA
Absolute amount of Des vaporized is much higher than other IA which would cause excessive cooling
Irritating to mucous membranes - LMA-may stimulate lead to laryngospasm d/t sleep and try to cough but not able & chords may spasm.
Traditional vaporizers could not compensate adequately
VP of 669 mmHg – near 1 atm
Traditional vaporizer would cause an anesthetic overdose
Des = 735 ml/min opposed to Iso = 47 ml/min
Differences btw measured flow vaporizers and Variable Bypass vaporizers
Measured Flow: Carrier gas (CG) bubbles through agent
Manual Temp. conversion
Multi-agent
Operator determines CG split
No longer manufactured (but still present on field machines)
Variable Bypass: CG flows over agent
Auto adjust for Temp & Pressure
Agent specific
Auto CG split
Higher safety standards
Temperature Regulation in vaporizers

Metals with high specific heat and high thermal conductivity are used to minimize heat loss
Close to wicks etc
Bypass chamber receives less flow with _______ temperatures
Bypass chamber receives more flow with _______ temperatures
Metals with high specific heat and high thermal conductivity are used to minimize heat loss
Close to wicks etc
Bypass chamber receives less flow with 'Lower' temperatures
Bypass chamber receives more flow with 'higher' temperatures
What is the effect of Temperature regulation thru the vaporizer?
↓Temperature → ↓Vapor Pressure
Sluggish/not hitting the sides of the vaporizer
More flow must go through the vaporizing chamber so that the patient receives the same concentration of agent
↑Temperature → ↑Vapor Pressure
Hitting the sides of the vaporizer rapidly
Less flow must go through the vaporizing chamber so that the patient receives the same concentration of agent
Effx of varied altitudes on Tec 6 Desflurane Vaporizer
Requires manual adjustments of the concentration dial with changes in barometric pressure
It is a pressurized vaporizer
It works at an absolute pressure
Partial pressure of Des changes with changes in ambient pressure
Required setting = Normal setting (%) X 760 mmHg/ambient pressure (mm Hg)
Tec 6 Carrier Gas composition effx on viscosity with gas other than O2
Decreased concentration output can occur with low flows of gases other than oxygen due to changes in viscosity
Vaporizer output is 20% less with nitrous than with oxygen
Nitrous oxide is less viscous than oxygen resulting in a lower working pressure at the differential pressure transducer
Take home – May have to increase dialed concentration with low flow anesthesia
Careful with low FIO2
Hazards with vaporizers
Incorrect agent
Simultaneous agent administration
Free Standing
Overfilling
Tipping
Vaporizer Safety Mechanisms
Push/Turn
Interlock-only one vaporizer at a time to be opened
Color Codes
“Keyed” filler system
How many PPM when you can smell the agent
5-300ppm
Oximetry
oximetry is produced by passing radiation through a sample and measuring the amount of absorbed radiation from a compound of interest. Msmt of a light source (radiation/heat/energy) absorbed by a substance.
Pulse Oximetry
Mandatory AANA/ASA standard monitor to aid in detecting hypoxemia
Noninvasive measurement of oxygen saturation (SpO2) in arterial blood
Sensor consists of 2 light emitting diodes (LED) and a light detector (a photocell/photodiode); a microprocessor analyzes the info and displays the SpO2
Placed across a pulsatile arterial bed (finger, toe, earlobe or any other perfused tissue)

combo of oximetry and plethysmography. Arterial pulsations are identified by plethysmography.