What is important to know in regards to specific entrainment ratio (Venturi)?
- the same FIO2 always has the same ratio
- altering flow rate will not alter ratio
How do you alter the FIO2 in a Venturi setup?
- by altering the port
What is important to know when using a venturi system?
- watch for bed sheet (covered port will increase FIO2 but decrease flow)
- always place collar to prevent blockage
What must be adjusted when switching a patient from a nasal cannula to a venturi setup?
- liter flow will need to be adjusted to insure correct FIO2
What is the FIO2 range for a venturi mask?
- 0.24-0.50
- high flow, but normally low concentrations
What is a mist tent (croupette)?
- high flow system
- plastic tent large enough to enclose child
- primarily used to provide aerosol therapy to children with croup
- air conditioned
What is the LPM & FIO2 delivered by a mist tent?
- 12-15 LPM to prevent CO2 buildup
- provides approximately 40% FIO2
What is an isolette ?
- high flow chamber designed to provide a thermally controlled O2 enriched, humid environment for newborns
- access provided through side ports to help maintain consistent FIO2
What is a head box?
- enclosure that fits over the head of a newborn
- warmed & humidified O2 is supplied to box by large bore tubing
What is the minimum flow rate for a head box?
- 7 LPM to flush out CO2
- higher flow may produce harmful noise
What must the flow be for an aerosol setup?
- 3 times the minute volume
What may be necessary for an aerosol setup at high FIO2's (greater than 0.60)?
- tandem setup may be necessary
What is another name for the "T" piece? What is it used for?
- Briggs adapter
- intubated patients & trached patients
What does an aerosol setup deliver?
- high humidity along w/ a precise FIO2