• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/127

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

127 Cards in this Set

  • Front
  • Back
What is the primary goal of mechanical ventilation?
to support the minute ventilation (VE) to meet the O2 and CO2 requirements for patients who cannot do so themselves.
What are the main parameters?
VT, RR (f), which combined make up the minute ventilation
What are the initial ventilator settings during volume ventilation what needs to be considered?
1. Minute ventilation (VE) = VT x RR (f)
2. Inspiratory gas flow
3. Flow waveform
4. Inspiratory and expiratory (I:E) ratio
5. Pressure limit
6. Inspiratory hold (pause)
7. Inspiratory and Expiratory pressure (PEEP)
What is the primary goal of VV?
to achieve a desired VE that matches the patient's metabolic needs and accomplishes adequate gas exchange.
How do you calculate VE?
Men VE = 4 x BSA
Women VE = 3.5 x BSA
How do you calculate IBW?
Women = 105 + 5(H-60)
Men = 106 + 6(H-60)
What are the normal Vt for adults and infants?
Vt = 6-12 ml/kg of IBW (adults)
Vt = 5-10 ml/kg of IBW (infants and children)
What are the recommended tidal volume for normal lungs?
10 to 12 ml/kg with RR 8 -12
What are the recommended tidal volume for COPD/Asthma?
8 to 10 ml/kg with RR 8 - 12
What are the recommended tidal volume for Restirctive Diseases?
4 to 8 ml/kg with RR 15 - 25

Remember: high rates may not provide enough time for exhalation resulting in intrinsic PEEP
What is the desired Plateau Pressure?
less than 30 cm H2O
Why are Vt's for more than 12 ml/kg not recommended?
because the risk of high pressures and accompanying overdistention and trauma to the lung as well as other complications.
Low volumes (4 to 8 ml/kg) for restrictive diseases prevent?
high pressures and alveolar distention
Low volumes less than 4ml/kg?
atelectasis
What is tubing compliance (CT)?
the amount of gas compressed in the ventilator circuit for every centimeter of water pressure generated by the vent during the inspiratory phase.
How do you calculate CT?
Change in pressure/change in pressure
What is compressible volume?
the volume of gas in the circuit
Define VDmech:
the volume of gas that is rebreathed during ventilation
What adds VDmech while ventilating a pt?
Flex tubing (6 inches)
Y-connector may add 75ml
HME may add 20 to 90ml
What reduces the VDmech while ventilating a pt?
ETT slightly reduces VDmech about 1ml/kg IBW
For some VV practitioners set?
f, Vt and flow
Some VV require what to be set?
rate, inspiratory time percent (Ti % of TCT), and VE
It is important to understand the interrelation of:
inspiratory flow, inspiratory time, expiratory tiem and I:E ratio
Formula for Total Cycle Time (TCT)
Ti + TE
Formula for rate
1min/TCT = 60sec/TCT (secs)
Formula for TCT
60sec/ f
Formula for I:E
Ti / TE
What is an inverse I:E ratio?
2:1 or 3:1. Expiratory time takes on the value of 1.
What are the complications of an inverse I:E ratio?
mean airway pressure
Formula for Ti?
Vt/flow
Ti can be determined when Vt and flow are known and the flow pattern is constant or square waveform.
Note: flow control calibrated in L/min so convert to L/sec.
Formual for Vt:
flow x Ti
Formula for flow:
f = Vt/Ti

Flow can be determined if Vt and Ti are known.
The clinician has the option to select flows and flow patterns on the ventilator.
In VV
The flow setting on a mechanical ventilator estimates the delivered flow of?
inspired gas
(Rate of gas flow)
High flow shorten Ti resulting in?
higher peak pressures adn poor gas distribution
Slow flow rates may:
reduce peak pressure
improve gas distribution
increase Paw at the expense of increasing Ti (result may cause cardiovascular side effects and shorten TE leading to air trapping)
Normal range for short Ti for normal lungs?
0.8 to 1.2 seconds
Normal I:E ratio?
1:2 or less (usually 1:4)
What is the initial peak flow?
about 60L/m (40-80L/min)
Longer Ti show to improve?
ventilation in nonhomogeneous lungs (ARDS) (taking 3 to 4 time constants)
Fast flow is beneficial for pt's with?
increased airway resistance (Raw)
i.e. COPD thus giving a longer TE and preventing or reducing air trapping
In selecting waveforms, the clinician must decide what?
if the Paw is more important for the patient than concerns of high peak inspiratory pressure.
True or False
High peak pressure may not always increase the risk of damage to the lung parenchyma.
True

(Pt's with acute asthma w/severe bronchospasm, mucosal edema, and increased secretion production)
What will changing the waveforms do?
can vary the peak flow delivered and the distribution of flow
True or False
In volume cycled machines, changes from constant to another flow pattern does change the peak flow selected.
False (does not)

It does change the Ti and the I:E ratio
In normal lungs, flow pattern is probably not of key importance. What flow patterns are?
descending and constant
Descending flow pattern may be beneficial by keeping peak pressures low and Paw high, and improving gas distribution for what types of pt's?
hypoxemic and low lung compliance patients
The descending pattern is more likely to deliver a set Vt at a lower pressure and provide for better distribution of air through the lung than a constant or an accelerating flow. What type of pt?
patients with high Taw
PaO2 decreases and PaCO2 increases while VE increases may be the result of?
auto PEEP, poor ventilation to perfusion matching adn changes in venous return.

This may require changing gas flow and/or pattern, Vt, f, or I:E ratio
What are used to determine Plateau pressure?
inflation hold or end-inspiratory pause
Inspriatory pause helps improve what?
the distribution of air throughout the lungs regardless of the type of flow pattern used, and to provide optimum ventilation/perfusion (V/Q) matching and reduce VD/Vt
Pressure targeted ventilation provides:
a set pressure to the patient during breath delivery while tidal volume can vary from breath to breath
Types of pressure ventilation:
PSV, PCV, BiPAP, dual control mode (pressure augmentation), VAPS, PRVC, VS
BiPAP
bilevel positive airway pressure
VAPS
volume assured pressure support
PRVC
pressure regulated volume control
VS
volume support
What is the advantage of PV?
provides flow on demand and potentially limiting pressures to avoid overinflation
In PV, initial flows may cause?
frictional forces (shearing) between adjacent alveoli with differing lung inflation characteristics (time constants)
Vt vary based on lung characteristics
To help preserve normal FRC an approptiate minimum level of PEEP would be?
3 to 5 cm H2O

Lack of physiologic PEEP may cause atelectasis
Low PEEP may be beneficial in patients with?
COPD who would normally pursed lip breath but cannot in an ETT
PSV goal if:
1. Help increase Vt (5 to 12ml/kg)
2. Decrease RR (less than 25 to 30bpm
3. Decrease WOB through artificial airway
A cycle criteria of 40% of peak low is an appropriate starting point for what type of pt's?
COPD patients
Initial settings for PCV are?
Rate, inspiratory time, and I:E ratio
the pressure gradient (PIP-EEP) is adjusted to establish volume delivery based on lung characterisitics and effort
PIP -5cmH2O as starting point
10 to 15 cmH2O
What type of waveform does PCV normally provide?
descending ramp waveform

adjust rise to meet pt needs
assess Ti
PCV is shown to improve what?
oxygenation and gas exchange, Paw, facilitate lung healing and reduce PIP, PEEP, VE, respiratory work, need for sedation, and ventilation time
What are the BiPAP initial settings?
5 to 10 cmH2O, increase in increments of 3 to 5 cmH20 until a rate of 25 bpm or lower is achieved adn Vt is 7ml/kg or more.
What are the EPAP initial settings?
2 to 5 cmH2O adn increased in increments of 3 to 5 cmH2O.
Initial settings of Dual Control Pressure Ventilation w/Volume Targeting?
provides the benefits similar to PV.
VAPS and PAug less familiar. Suggested initial settings.
PS with a descending flow pattern may be implemented.
Volume mode must be selected VC-CMV (assist) or VC-SIMV.
Initial settings for VAPS or Paug?
Target volume selected based on VV criteria.
Sensitivity set (flow or pressure) so that patient can trigger a breath.
Set upper pressure limit (max safety limit), i.e. 40 cmH2O
PIP for breath delivery is set to provide the majority of the Vt (set P = Pplateau - PEEP)
Baseline pressure set the same as VV
Activate Paug (or VAPS)
Set the flow waveform on rectangular and set inspriatory gas flow setting so that the pressure/time and flow/ time graphics produce the desired effect.
Flow pattern and flow delivery important in Paug. What is the criteria for setting flow?
1. flow must be set hight enough so tha tTi does not last longer than Ti for a mandatory, unassisted volume breath.
2. Flow must be set so that the pressure waveform during a Paug breath is rectangular at the beginning of inspiration and has a small pressure rise at the end of inspiration for some breaths.
3. the rectangular flow waveform should be selected becasue use of the descending ramp flow reduces some of the advantageous effects of the augmented breath and prolongs inspiration.
Initial settings of Pressure Regulated Volume Control (PRVC)?
provides closed loop pressure breaths and targets the pressure to achieve the set volume.
it is pressure limited, time cycled mode that uses the set Vt as a feedback control.
the operator sets a Vt to be delivered that is appropriate for the patient. Baseline pressure and pressure limit are also set.
Ventilator delivers one or more test breaths. (sometimes pressure targeted breath or volume targeted breath)
Ventilator calculates Cs adn Raw in order to determine the pressure required to achieve the set Vt.
Ventilator progressively adjusts the pressure level unitl set Vt is achieved.
Ventilator will stop pressure delivery at 5 cmH2O below max pressure limit.
Why is it important to set an upper pressure limit in PRVC?
1. provides the upper limit for the ventilator.
2. if patietn coughs or forcibly exhales during the inspiratory phase, the ventilator will not permit pressure to exceed the upper pressure limit.
Initial settings of Volume Support?
purely a spontaneous mode.
operator set sentitivity, Vt, and upper pressure limit. So Vt based on VV criteria. (Set Vt is the min. Vt)
less pressure required as lungs improve
Backup mode for apnea
Mandatory Minute Ventilation
operator set the minimum parameters (Vt, f, and high and low pressure alarms)
ventilator adjust according to patient needs.
Selection of FiO2
goal is to achieve acceptable PaO2 between 60 to 100mmHg
What is the formula for estimating desired FiO2?
PaO2(desired) x FiO2(known)
-------------------------------------------
PaO2(known)
Sensitivity
-set so that patients can easily flow or pressure trigger a breath.
-Flow triggering: 1 to 10L/min below base flow. (manufacturer/operator set)
-Pressure sentitivity: -1 to -2 cmH2O
-Flow triggering now the preferred method for triggering because it has a faster response time.
Humidification
-Normal spont. breathing provide 100% RH at 37 degrees centigrade adn contains 44 mg/L of water.
-Conditioning typically occurs down to the fourth or fifth genteration of subsegmental bronchi and is called the Isothermic saturation boundary.
Types of Humidifiers:
1. Passover
2. Vapor phase
3. Wick
4. Active Heat adn Moisture Exchange
Heat Moisture Exchangers (HME)
artificial nose
can add 10 to 14mg/L of water
hygroscopic HME (HHME): 22 to 34mg/L
resistance to flow (accumulates moisture and secretions) in extended use.
HME add mechanical deadspace (50-100ml)
change every 2-3days
watch for thickening secretions
ET occulusion in greater than 7days.
Alarms
-low pressure 5-10cmH2O below PIP
-High pressure 10cmH2O above PIP
low PEEP/CPAP 2 to 5 cmH2O below set PEEP/CPAP
apnea alarm: set no mor ethan 20 sec.
ratio alarm: warns when Ti is mor than TCT
low Vt: 10 to 15% below set Vt
low exhaled VE: 10 to 15% below average VE
FiO2 5% above and below set FiO2
Action during Ventilator Alarm Situations
1. ensure pt is being ventilated
2. when in doubt disconnect pt and manually ventilate
3. cannot correct problem, replace ventilator
4. operating manual may assist with troubleshooting
5. final resort-call local company rep.
Ventilation Setup
check/calibrate ventilator
fill humidifier
adjust alarms
verify monitoring
emergency equipment in close proximity
check suctioning equipment
Initial Ventilator Settings for COPD
-non-invasive ventilation is possible
-VC or PC-CMV to relieve increased WOB
-Adjust peak flow to meet pt's demand
-Vt 8 to 10ml/kg with RR 8 to 12bpm
Ti 0.6 to 1.2sec
PEEP approx. 5 cmH2O
-Monitor hyperinflation (auto-peep) by setting the lowest possible VE that produces acceptable gas exchange (Pt. baseline PaCO2)
-Provide the longest expiratory time possible
-if pt. initiating inspiration, and auto-peep noted, set PEEP near 80% of the auto-peep level (3-5cmH2O)
-Plateau pressure should be monitored and maintained less than or equal to 30cmH2O to avoid alveolar over distention. (May req. sedation)
-PaO2 at 55 to 75mmHg
Neuromuscular disorders
-full or partial support
-negative or PPV
-noninvasive or invasive ventilation
-Assist/Control mode (CMV)
-Volume ventilation
-High Vt (12 to 15ml/kg)
F = 8-12bpm
-Inspriatory flow rates approx. 60 lpm (1sec Ti)
-Constant or descending waveform
-PEEP 5cmH2O/ FiO2 21%
Guidelines for Asthma
-VC or PC-CMV
-keep peak and plateau pressures at min.
-FiO2 to keepp PaO2 600-100mmHg
-Permissive hypercapnia (PaCO2 45-80mmHg)
-Possibly use sedation and paralytic
-PEEP to apporx. 80% of auto PEEP
-Long TE: RR 8 or less, Vt 4 to 8ml/kg; Ti less than 1 sec.; peak flow 80 to 100 lpm with descending flow waveform
Guidelines for Closed Head Injury
-Protect airway due to ALOC
-V-CMV or P-CMV (watch pressures)
-Monitor ICP and Hypoxemia
-Vt 8 to 12ml/kg; plateau pressure about 30cmH2O
-RR 15 to 20bpm (avoid auto-peep)
-initial FiO2 at 100. Titrate to keep PaO2 70 to 100mmHg. Ti 1sec/descending waveform
-PEEP 0-5cmH2O
Adult Respiratory Distress Syndrome
-PC or VC-CMV
-SaO2 88 to 90%. FiO2 titrated/begin @100%
-May need to sedate or paralyze
- High PEEP (15cmH2O or greater)
-Plateau pressure less than 30cmH2)
-Permissive hypercapnia
Guidelines for CHF
-Select mode that reduces WOB
-May use non-invasive mask for NPPV or CPAP to improve oxygenation.
-Consideration of PPV and hemodynamics
-VC or PC-CMV recommended
-Vt 8-10ml/kg; f 10 or above; Peak flow 60 lpm; Ti 1 -1.5sec
- PEEP 5 to 10cmH2O
- FiO2 100 then titrate to keep SpO2 90-92%
-Monitor ABG's, urine output, SpO2, Hemos
Patient Evaluation
-observation by well trained, caring clinician
-check physician orders
-operational verification procedure (OVP)
-overall check of equipement (vent, circuit, humidifier)
AARC guideline for documentation of Pt ventilator system:
1. Data recorded on appropriate form
2. Pt/vent system check
3. Physician orders
4. Brief narrative of clinical observation
First thirty minutes
-assess B/S
-check vitals
-check alarms
-ABG
-chest xray
-ventilator check
Documentation Guidelines
-mode
-sensitivity
-Vt, rate, VE
- corrected tubing compliance
-alveolar ventilation
-dead space
-mechanical deadspace (Added)
-final alveolar ventilation VA = (Vt-VDant-Added VDmech) x f
Monitoring Airway Pressure
Peak inspiratory pressure (PIP
Plateau pressure
Set pressure (PC)
PIP-Pplateau= Transair press
End expiratory pressure
Mean airway pressure
Pressure Limit
Low pressure alarm
Circuit assessment for leaks
Vital Signs, Blood Pressure, and Physical Exam of the Chest
-ECG-noninvasive means of continuously monitoring HR and rhythm
-Temp
-Systemic arterial blood pressure
-Central venous pressure
-Pulmonary artery pressure
-Chest exam (inspection, palpation, percussion, and auscultation)
Management of ET and Tracheostomy Tube Cuffs
check every 8-12hrs
keep below 20 to 25mmHg/ 27-34cmH2O to reduce risk fo tracheal damage.
Use MLT whenever possible
Cut in the Pilot Tube
Accidental cuts/punctures are common occurances
1. Manufacturer designed blunt tipped needle device used as one solution
2. Stop cock method
3. Exchange ET tube
Tube and Mouth Care
ETT respositioned every 8 to 12hrs
Mouth care done routinely (VAP's)
Monitoring Compliance and Airway Resistance
-Normal Cs: 70 to 100ml/cmH2)
-Assess for changes in lung condition
-CD decreases whenever Cs decreases or Raw increases.
-Raw 0.6-2.4cmH2O/L/sec. Treatment of Raw must be directed at the specific cause of increase (i.e. sx, clearing obstruction, or bronchodilator)
Bedside Measurement of Pressure Volume Curves
daily evaluation of the pressure volume relationship (either static or dynamic) is recommended to tract progression of lung and airway changes.
Comment Section of the Venitlator Flow Sheet
-Pt assessment (B/S, skin color, LOC)
-Changes in ventilator settings
-Changes in physician's orders
-Description of any equipment problems
-STOP (Subjective, Therapeutic Objective, and Plan)
Ventilator Graphics
-Provide immediate visual infromation about the pt-vent. interaction
- Provide a graphic record of pathophysiological changes
-Graphics assists in monitoring the functioning of the ventilator and evaluation of the pt's response to the ventilator and to help the clinician adjust the ventilator settings
Scalar:
used to specify the waveforms for pressure, flow, and volume that are graphed against time (i.e. pressure, flow and vol scalars)
Loop:
describes the type of graph (displays of 2 variables). Ex: pressure-vol adn flow-vol loops
Name the waveforms for pressure:
rectangular, exponential (rise)
Name the waveforms for volume:
Ascending ramp, sinusoidal
Name the waveforms for flow:
rectangular, sinusoidal, ascending ramp, descending ramp, exponential (decay)
The formula for resistance:
change in pressure/flow
Monitoring Data
-Flow and pressure monitors on the inspiratory side of the ventilator just before the point where gas exits the ventilator and goes to the patient.
-Flow and pressure monitors on the expiratory side of the ventilator just as gas enters the ventilator through the exhalationi valve.
Relationship of Pressure, Volume, Flow and Time
1. Volume = Flow x Ti
2. The flow of gas into the lungs depends on the difference between the pressure from the power source (the ventilator) and the pressure inside the lungs. (Flow = V/Ti)
3. The amount of pressure required to inflate the lungs depends on lung compliance and airway resistance.
Volume Ventilation (Constant Flow)
-Force required to deliver air to lungs with or without decreased compliance or increased airway resistance.
- High internal resistor: placed in-line between the gas source and lungs to regulate the amount of flow.
- High pressure source prevents back pressure.
Auto PEEP
-Expiratory pause allows direct measurement of auto-peep
-Some ventilators give total which is the sum if auto-peep and set peep
Characteristics of VV with any Fixed Flow Pattern
- the flow waveform provided, regardless of type, does not change shape during inspiration even with changes in lung characteristics.
-the flow waveform remains the same from breath to breath
- volume delivery remains constants as long as the breath is volume cycled.
-pressure patterns at the upper airway vary with changes in CL and Raw and the type of flow pattern selected.
Pressure Ventilation
-(PC-CMV) the operator sets a pressure to be delivered to the pt's lungs. Flow and Vt vary.
-Ventilator attmepts to provide a set pressure from start of inspiratory until expiration begins producing a nearly constant pressure waveform.
PV with Constant Pressure Waveforms
4 important characteristics:
1. rectangular (constant)
2. pressure waveform not affected by changes in lung characteristics or pt flow demand
3. the rate of flow delivery varies according to the lung characteristics, set pressure, and inspiratory effort.
4. the flow waveform rises rapidly at the beginning of inspiration and decreases during inspiration (continuously variable decelerating pattern)
Pressure Ventilation vs. Volume Ventilation
VV: constant flow (rectangular pattern)
PV: flow waveform is a descending curve which vaires with lung characteristics and patient flow demand. (continuously variable decelerating waveforms)
VV: Pressure curve resembles ascending ramp or rising exponential curve.
PV: pressure curve rectangular (Ti long enough)
VV: the pressure waveform changes with changes in lung characteristics (Cs, and Raw)
PV: changes in Cs and Raw change flow and volume delivery but the pressure waveform remains the same.
Inspiratory Rist Time Control: Sloping or Ramping
-pressure breath produce high flow at the beginning of inspiration.
-small ET tubes or Raw create turbulent flow
-pressure spike can occur at the beginning of the pressure curve before the pressure adjusts to the set value
-flow and pressure delivery can be tapered at the start of inspiration adn thus reduce the spike.
-insp. flow delivery during PCV, can be adjusted with inspiratory rise time control aka slope control.
-slope/rise time function controls the rate at which the flow valve opens.
Rise/Slope Function
-adjusting inspiratory rise slows or speeds the rate at which pressure and flow exit the ventilator for a specific period.
-the adjustment may make the breath more comfortable
-matches ventilatory breath delivery with pt flow demand thus improve pt ventilator synchrony and decrease WOB
Rise Time/Slope Function Caution
too much tapering of flow and pressure delivery can reduce Vt and cause pt ventilator dyssynchrony
Pressure Support Ventilation
-is a pt triggered, pressure targeted, flow cycled mode of ventilation.
-as lung characteristics deteriorate, pressure remains constant/volume delivered decreases.
-operator sets pressure, sensitivity, inspiratory rise time (slope), and inspiratory flow cycle value, (aka flow cycle pattern)
Flow Cycling During PSV
-flow cycling occurs when the ventilator detects a decreasing flow which represents the end of inspiration.
-appropriately set flow cycling percent varies with pt's
-COPD pt's or pt's with increased Raw may require higher flow cycle percents (40%)
Automatic Adjustment of Flow Cycle Criterion
-difficult to adjust flow cycle criterion due to frequently changing lung mechanics (Cs and Raw)
-software program available in some ventilators (unavail. in the US)
Pressure Volume Loop
-another graphic available on newer vents.
-pressrue and volume determine compliance and continuous gas flow during the curves also provide information about Raw.
Spontaneous Breaths and P-V Loops
-a practitioner can learn to distinguish mandatory breaths from spontaneous breaths by observing the way the P-V loop is created during breath delivery.
-spont.: trace clockwise
-positive pressure breath: counter clockwise tracing
Flow Volume Loops during Mechanical Ventilation
-F-V loops used for various purposes.
-inspiration is above the baseline
-exhalation is below the baseline
-Peak Expiratory Flow Rate (PEFR) is one of the parameters measured on the F-V loop. It is the highest value on the expiratory flow curve.
Troubleshooting F-V Loops During Mechanical Ventilation
-the F-V loop can be used to detect leaks and air trapping (auto-PEEP)