Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
59 Cards in this Set
- Front
- Back
Which one of the 3 types of quantitative analysis errors is the following:
easy to recognize since they involve a major breakdown in the analytical process, such as samples being spilt, wrong dilutions being prepared, or instruments breaking down or being used in the wrong way. |
Gross Error
|
|
Which one of the 3 types of quantitative analysis errors is the following:
characterized by precise and inaccurate measurement results. |
Systemic error
|
|
Which one of the 3 types of quantitative analysis errors is the following:
characterized by widely scattered, i.e., imprecise, measurement results. |
Random error
|
|
Accuracy and precision:
the calculated value for __ provides a formal expression of the scatter in the results from an analysis rather than the visual judgment shown previously. |
standard deviation
|
|
provides an exact description of how the analysis is carried out.
|
Analytical procedure
|
|
expresses the closeness of agreement(degree of scatter) b/w a series of measurements obtained from multiple sampling of the same homogeneous sample
|
precision
|
|
precision obtained under the same operating condition over a short period of time(an assay is repeated by the same person using the same equipment)
|
Accuracy
|
|
expresses the within-laboratory variation of precision when the analysis is carried out by different analysts, on different days, and with different equipment
|
intermediate precision
|
|
expresses the precision b/w laboratories such as when a method is transferred from one part of a company to another.
|
reproducibility
|
|
simplest method is to compare a substance being analysed with a reference standard analysed by the same procedure (the reference standard is a highly characterized form of the drug that has been subjected to extensive analysis)
|
accuracy
|
|
relates to the performance of the analytical equipment(the four Q's)
|
system suitability- these test are designed to show that the system meets the standards required by the method
|
|
consists of all of the reagents or solvents used in an analysis without any of the analyte being present
|
analytical blank
|
|
comparison of the value or values of a particular parameter measured by the system under strictly defined conditions with the preset standard values
|
calibration
|
|
the smallest amt of an analyte which can be detected by a particular method.
|
limit of detection
|
|
smallest amount of analyte which can be quantified reliably
|
limit of qualification
|
|
most analytical methods are based on processes where the method produces a response that is linear and which increases or decreases linearly with analyte concentration
|
linearity
|
|
refers to the interval between the upper and lower concentration of an analyte for which an acceptable level of precision and accuracy has been established
|
range
|
|
evaluated to determine how resistant the precision and accuracy of an assay are to small variations in the method (for example, the stability of analytical solutions, or the length of sample extraction time)
|
robustness
|
|
measure of how capable a method is of measuring the analyte alone in the presence of other compounds contained in the sample
|
selectivity
|
|
indicates how responsive the method is to a small change in the concentration of an analyte (may be viewed as the slope of a linear response curve)
|
sensitivity
|
|
indicates how responsive the method is to a small change in the concentration of an analyte (may be viewed as the slope of a linear response curve)
|
weighing by difference
|
|
often encountered in relation to the composition of mobile phases used in high-pressure liquid chromatography (for example, when 30 ml methanol is mixed with 70 ml water, a 30:70 v/v mixture is formed)
|
v/v
|
|
usually used to express the content of an active ingredient in liquid formulations such as injections, infusions, and eyedrops
|
w/v
|
|
a common measure used to express the concentration of active ingredient in a formulation such as a cream
|
w/w
|
|
common measure used for impurities in drug substances, particularly heavy metals and solvents
|
ppm
|
|
physical properties of drug molecules, such as ________________
__________, along with simple chemical ___________ and degradation reactions, play an important part in the design of analytical methods |
pKa + partition coefficient
derivitization |
|
what is the most effective range for a buffer
|
1 pH unit either side of the pKa value of the weak acid or base used in the buffer
|
|
the partition coefficient for a compound (P) can be simply defined as ________ where Cois the concentration of substance in an organic phase and Cwis the concentration of substance in water
|
P=Co/Cw
|
|
the ________ the more a substance has an affinity for organic media
|
Greater the P
|
|
in zero-order kinetics the rate of degradation is WHAT in regards to reactant concentration?
|
Independent of the concentration of the reactants.
|
|
1st order degradation has what typical degradation pattern?
|
The hydrolysis of a drug in solution
|
|
The lack of what about the double bond ensures that the stereochemistry of a drug and its isomer are different?
|
lack of free rotation
|
|
when is a compound said to be chiral
|
when there is no symmetry around carbon
|
|
when __ is present in a moecule, there is a possibility of diastereomers
|
more than one chiral center
|
|
analytical derivitization is employed for two reasons:
to improve the analysis by improving the __ and to improve the analysis by improving chromatographic behavior of detectability, i.e. __ |
selectivity
sensitivity |
|
Ultraviolet and Visible Spectroscopy:
radiation in the wavelength range _________________ is passed through a solution of drug |
200 to 700nm
|
|
most drug molecules absorb radiation in the ultraviolet region of the spectrum, i.e., ___________
|
100 to 350nm
|
|
short wavelength UV radiation ________ can cause the strongest bonds in organic molecules to break
– weaker bonds in molecules are of more interest to the analyst because they can be excited by longer wavelength UV radiation __ _______ which is at a longer wavelength than the region in which air (for example, oxygen and carbon dioxide) and common solvents absorb |
<150nm
200nm |
|
chemical substructures that are responsible for the absorption of electromagnetic radiation are called ______________(1)
–a most common ___________(1) in drug molecules is the ________ __________________________(2) |
(1) chromophores
(2) aromatic ring, such as the benzene ring |
|
the___________ by a solution of molecules is governed by the Beer-Lambert Law
|
measurement of light absorption
|
|
–
in pharmaceutical products, concentrations and amounts are usually expressed in grams or milligrams, and, for the purposes of analysis of these products, the Beer-Lambert equation is written in the following form, _________________ |
A=A(1%,1cm)bc
|
|
Ultraviolet and Visible Spectroscopy:
the __________, a deuterium lamp for the UV region from 190 to 350 nM and a quartz halogen or tungsten lamp for the visible region from 350 to 900 NM |
light source
|
|
since measurements are usually made in a 1 cm cell, the equation can be written as _________________ which gives the concentration of the analyte in g/100 ml
|
c=(a/A(1%, 1cm)
|
|
the ______________, used to disperse the light into its constituent wavelengths
|
monochromator
|
|
the _____, may be designed to split the light beam so that the beam passes through two sample compartments, and, in such a double beam instrument, a blanksolution can then be used in one compartment to correct the reading or spectrum of the sample
|
optics
|
|
____ are generated in a high-vacuum region, or immediately prior to a sample entering a high-vacuum region, using a variety of methods for ion production
|
charged molecules of molecular fragments
|
|
GC/MS and LC/MS provide highly sensitive and specific methods for determining drugs and their ____________________________
|
metabolites in biological fluids and tissues
|
|
which of the two is cheaper and more sensitive?
Mass spectrometer or magnetic sector instrument |
Mass spectrometer
|
|
in this form of analysis, spectra may contain molecular ion only
|
Mass spectrometry
|
|
the most frequently used analysis technique in pharmaceutical analysis
|
chromatography
|
|
the length of time it takes a retained compound to pass through the column depends on the compounds ________________ which is a measure of the degree to which the compound partitions (adsorbs) into the stationary phase from the mobile phase
|
capacity factor
|
|
what would happen in chromatography if a compound does not partition appreciable into stationary phase.
|
It will travel through the column at the same rate as the solvent
|
|
Calculation of column efficiency
the ______ a chromatographic peak is relative to its retention time, the less efficient the column it is eluting from |
broader
|
|
•
Calculation of column efficiency – the efficiency of the column is most readily assessed from the width of the peak at its height W1/2 and its retention time using the following formula, n= 5.54(tr/W1/2)2 – where, nis the number of _______________, where, tris the time taken for the compound to pass through the column, where, W1/2is the width of the peak at half its height – _______________ is usually expressed in theoretical plates per meter, or nx (100 cm/m)/L – where, Lis the column length in cm |
theoretical plates
column efficiency |
|
•Stationary and mobile phases
–there are two principal mechanisms which produce retention of a compound passing through an HPLC column –silica gel, a ___________ column packing, where the mechanism of retaining a compound is by ____________________ onto the polar groups of the stationary phase –octadecylsilane-coated (ODS-coated) silica gel, a ____________column packing, where the mechanism of retaining a compound is due to _______into the stationary phase. |
-Normal-phase
-adsorption of the polar groups of a molecule -reverse plase -partitioning of the lipophilic portion of a molecule |
|
High-pressure liquid chromatography (HPLC)
•Structural factors which govern the rate of elution of compounds from HPLC columns –neutral compounds –for a neutral compounds it is the balance between its __________ __________ which will determine the time it takes to elute from an HPLC column –the pH of the mobile phase does not play a part –in the case of a reverse-phase column (ODS-coated silica gel), the _____________ a compound is the more it will be retained –in the case of a normal-phase column (silica gel), the _________ a compound is the more it will be retained –polarity can often be related to the number and strength of hydrogen-bonding functional groups in a molecule |
-polarity and lipophilicity
-more lipophilic -more polar |
|
HPLC:
– the control of the rate of elution via control of the pH of the mobile phase is only applicable to compounds in which the _____________________ (the majority of common drugs) |
degree of ionization is dependent on pH
|
|
reverse-phase chromatography increases the
|
retention time
|
|
•Summary of detectors used in HPLC
–for the majority of analyses of drugs in formulations, variable wavelength ___________ are used which are based on the absorption of UV light by an analyte –the diode array detector ( UVDAD) is an advanced type of UV detector with the ability to monitor across the full UV range simultaneously –________________________ (ELSD) which is based on the scattering of a beam of light by particles of compound remaining after evaporation of the mobile phase |
-UV detectors
-Evaporative light scattering detector |