10 Cards in this Set
- Front
- Back
What protects the scavenger hose from being connected to something else?
|
DISS
|
|
Is the scavenger hose connected to regular hospital suction?
|
No, must be seperate (regular suction air is recirculated w/in the hospital)
|
|
What is the minimum suction for a scavenger (at the wall)?
|
30LPM
|
|
What are the OSHA requirements for halogenated agents alone?
|
2 ppm
|
|
What are the OSHA requirements for N2O alone?
|
25 ppm
|
|
What is the ANSI standard leak rate for ventilators?
|
Less than 100 ml/min resulting in < 3.7 ppm ambient concentraiton of N2O
|
|
How large is the reservoir bag on a closed system?
|
3-5L
|
|
What is the output/positive pressure relief valve of a closed scavenger system?
|
Opens if gas flow > capacity of the vacuum system, which vents excess gas into the room
|
|
What is the Subatmosphereic pressure relief valve/intake valve of a closed scavenger system?
|
Opens and entrains room air if waste gas flow to scavenger < vacuum capacity -> prevents loss of gas from the breathing system.
|
|
What are the OSHA Requirements for N2O/Hallogenated?
|
0.5/25 ppm
|