What is the transmission formula? |
HVL = 1/2ⁿ |
|
|
What is the field size on the skin (surface) formula?
|
SAD - depth/ SAD |
|
|
Transmission formula = attenuation
|
It = Iₒ e- 0.693 x thickness/ HVL |
|
|
TVL formula |
TVL = 3.32 x HVL
|
|
|
Linear coefficient |
ɥ = 0.693 / HVL cm-1
|
|
|
Given dose (or dmax dose) |
(tumor dose/Pdd) X 100
|
|
|
TPC
|
(273.2 + Cᵒ/ 295.2)(760/P)
760 is in mmgh use 1013.2 mbar |
|
|
Celsius conversion to farenheit
|
C = 5/9 (F - 32) |
|
|
Mayneord Factor (Big F)
|
(New + dmax/ Old SSD + dmax) X (Old SSD + depth/ New SSd + depth) |
|
|
Decay formula
|
A = Aₒ e - 0.693 x time /T 1/2
|
|
|
Penumbra
|
GP = s(SSD + depth -SDD)/SDD |
|
|
Magnification |
Img distance/ Object distance
or Img size/ Object size |
|
|
Exposure to dose conversion
|
Exposure x dose = rads |
|
|
Dose to exposure conversion |
Dose / exposure = R
|
|
|
Dmax or Given dose
|
(tumor dose / pdd)x 100 |
|
|
Inverse Square
|
I₁D ₁²= I₂D₂² |
|
|
Scatter to Air Ratio = SAR |
Scattered dose at given point in phantom / dose in free space at same point |
|
|
Equivalnt Square = EQSQ
|
2A / P |
|
|
Wedge factor |
dose with wedge / dose without the wedge |
|
|
RX given dose or dose to a point |
RX dose @ depth / Pdd @ RX dose |
|
|
Equivalent Radius = EQRA |
2A / square root of pi *P |
|
|
# e- per gram formula |
NA x Z / A |
|
|
TAR
|
Dose in the phantom / dose in air at the same point |
|
|
Pdd
|
Depth / dmax X 100
|
|
|
Electron density formula
|
Pe = Pm Na Z /A |
|
|
Avogados number (number of atoms)
|
Na = 6.02205 X 10²³ atoms |
|
|
Atoms per gram formula
|
NA / A |
|
|
As Pdd increases
|
E↑ Field Size↑ depth↓ SSD↑
|
|
|
As TAR increases
|
E↑ Field Size↑ depth↓
Independent of SSD or SAD |
|
|
As TMR increases
|
E↑ Field Size↑ depth↓
Independent of SSD or SAD |
|
|
Wedge factor is dependant on
|
Energy, Field size, depth
|
|
|
GTV
CTV PTV |
Gross tumor volume, clinical tumor volume, planning target volume |
|
|
Which one of these is most important .
GTV CTV PTV |
PTV |
|
|
Back scatter factor = BSF
|
dose at dmax / dose in air |
|
|
OD = optical density
|
log lₒ / lᵻ |
|
|
GAP formula |
(L₁x d /2SSD₁) + (L₂x d/2SSD₂)
|
|
|
Collimator formula |
(1/2 L₁x 1/SSD) -1 tan |
|
|
Couch formula |
(1/2 L₁x 1/SAD) -1 tan |
|
|
TPR = tissue phantom ratio
|
dose in tissue at a point / dose in phantom at referenced depth |
|
|
Atomic # Z what is the division between #protons and e- and neutrons
|
number of protons + number of e- = Z
|
|
|
Atomic mass A
|
number of protons + # neutrons = A atomic mass
|
|
|
What is 2n²?
|
n is the quantum number |
|
|
MU formula
|
TD/Junk
|
|
|
T or F |
True |
|
|
Calculate the MU of a 6 MV e- beam with a 15 x 15 cm cone. RX dose is 200 cGy at 95 % isodose line. The output of a 15 x 15 cm cone is 1.042. |
MU =TD/Junk
OR
SAD Technique {MU=cGy / C F×(OF×TPR×WF)} |
|
|
Calculate the MU of a 6 MV e- beam with a 15 x 15 cm cone. RX dose is 200 cGy at 95 % isodose line. The output of a 15 x 15 cm cone is 1.042. If there is a 1cm bolus what is MU? |
MU =TD/Junk
Fixed SAD Technique MU=cGy / CF×(OF×TPR×WF)
Fixed SSD Technique MU=cGy.SSD / CF×(OF×PDD×WF) |
|
|
Rule of thumb |
E/2 |
|
|
T or F |
True |
|
|
Compton and photoelectric interactions are produced by what machine energy ranges |
Compton by Mega (bone- high Z) |
|
|
T or F you can use MLCs with Island blocks?
|
False you cannot
|
|
|
Calculate the gap of 2 adjacent fields which are |
gap = L₁*d/2SSD + L₂*d/2SSD
10*7/200 + 15*7/200 = |
|
|
%dd for a 10 x 10 field at 10cm with 80cm SSd (dmax at 1cm) using a 4 MV photon is 62.2. Find the %dd for the same field size and depth at 100 cm |
(NEW SSD +dmax/OLD SSD +dmax)²*(OLD SSD + depth/ NEW SSD +depth)² x original pdd |
|
|
What is the collimator angle if the spine field is 18 cm long at 80 SSD?
|
(1/2 L* 1/ssd) arc tan -1
(18/2 * 1/80 )*tan -1 = 6.4 |
|
|
What is the couch angle if the brain field is 16 cm in length at 100 SAD?
|
(1/2 L* 1/sad) arc tan -1 |
|
|
T or F
For orthovoltage (E up to 400 kvp) and lower E xrays the dmax is on the skin and always 100%? |
True because dmax is always on the skin
|
|
|
Calculate the Hvl of a material if the thickness to produce 60% transmission is 7cm. |
It = I₀ e- 0.693*t / HVL |
|
|
During a 2.5 minute fluoroscopy exam a tech was exposed to 250 mR/hr. What is the exposure? |
Exposure = Rate x Time |
|
|
What is Tar at 5 cm depth if dose at 5cm is 180 cGy and dose in freespace is 210 cGy?
|
180cGy / 210 cGy = .857 or .86
|
|
|
Definition |
HVL |
|
|
Definition |
TVL |
|
|
Hinge Angle |
180/2 - wedge angle |
|
|
Wedge Angle |
90- hinge angle / 2
|
|
|
Tray factor |
Dose with tray / dose without the tray
|
|
|
Exposure
|
exposure rate X time
|
|
|
Radium Equivalent - brachy
|
『A/『ra
OR source * activity / Ra source |
|
|
Exposue at distance - brachy
|
『A /d² |
|
|
Exposure time - brachy
|
『 A t/『ra
|
|
|
Mghr -- perm implant
|
RaEq *1.44 * T1/2 *24 hrs
|
|
|
Mghr -- temp implant
|
RaEq * T1/2 * 24 hrs
|
|
|
Field weighting |
When beams are weight equally each field receives the same amount of dose --180/4 = 45 cGy per field
As weight is increased the hot spot moves toward the surface of that beam AP/PA field weighted 3:1 with a dose of 200 200/4 = 50 3 * 50 = 150 1 * 50 = 50 |
|
|
What is image size of a block if magnification is 3.2 cm and object size is 1.25?
|
mag = img size/obj size
3.2cm = image size/ 1.25 3.2 * 1.25 = 4 |
|
|
What is the depth of 50% of a 10 MeV e- energy?
|
E/2 = 10/2 = 5
???? |
|
|
The atom is neutral provided the # of e- and the # of protons is the same. If the atom has more e-, is the atom neg or positive? |
more e- = neg atom b/cuz e- are neg |
|
|
AMU
|
atomic mass unit |
|
|
Isotopes (ZA)
|
Nuclear with the same # of protons and different # of neutrons |
q
|
|
Isobar (AZ)
|
have the same atomic weight A different Z
do not have same chemical reaction |
|
|
Isotone (N) |
Isotone have the same # neutrons different Z and different A
|
|
|
Annealing |
heating to remove radiation from ionization chambers
|
|
|
How is absolute dose measured? |
Ion chambers and Fricke dosimetry.
absolite dose means 1 rad is 1 rad |
|
|
Who calibrates ion chambers and how often
|
Calibrated by ADCL or NIST |
|