28 Cards in this Set

  • Front
  • Back
True or False
If input one is more negative than input 2 the deflection of the waveform will be up
True
Impedance is defined as:
the resistance against alternating current flow
A bipolar montage follows which rules?
combines serial pairs of electrodes into a chain, in a chain of electrodes input 2 of one channel is the same electrode as input 1 of the next, left side electrodes appear first followed by the right side
What is the amplitude (uV) of activity that measured 12 mm in height, if a 50 uV calibration signal measured 6 mm?
100 uV
nasopharyngeal electrodes are placed where:
through the nasal passage
If the vertical scale (sensitivity) is 5 uV/mm, then a 60uV will measure:
12 mm
(60uV/5uV/mm)
Best montage to measure the amplitude of a sharp wave?
A2 ear reference montage
A faster sweep speed (60mm/sec) allows:
easier measurements of faster frequencies, easier ID of 60 hz artifact, better ID of degree of synchrony between 2 hemispheres.
The use of 60hz notch filter does what:
eliminates 60 hz activity only
The average reference recording is best used for:
localize focal activity
Sensitivity is conventionally defined by:
ratio of voltage to height
What type of montage combines activity from many electrodes to form a reference?
Average reference
which type of montage most gives the true amplitude EEG activity at a given electrode?
non-cephalic reference
The time constant of .1 second is aprox. equivalent to what low filter setting?
1 Hz
Common mode rejection ratio is closely related to ?
A balanced amplifier
what is the least invasive electrode application?
subdermal
A type of amplifier that amplifies the differences in voltage between 2 inputs?
differential amplifier
electrode "pop" is induce by what?
poorly chlorided electrodes
The time constant is determined by?
resistance x capacitance
The laplacian montage is often used for?
topographic EEG mapping
time scale of 3 cm/sec, what is the duration in mm for a 10 hz wave?
3 mm
The average reference is created by :
averaging the activity from all or (most) electrodes used.
High pas filter is:
eqivelant to low frequency filter
bipolar montage is best used to:
localize focal activity
True or false
The bio cal demonstrates the integrity of the system from patient to instrument.
True
Transverse bipolar monitoring is more accurate than longitudinal montage in localizing the activity:
spreading from temporal to parasagittal region.
What is the duration of a wave measuing 5 mm, when the 20 mm calibration signal is equal to 200 msec?
50 msec
(10 x 5mm = 50)
Which seizure type (ictal event) least likely shows abnormality?
?