88 Cards in this Set

  • Front
  • Back
the doppler effect occurs
all echoes from moving sources
the doppler effect depends on
the closing velocity between transducer and tissue, carrier frequency, and u/s propagation velocity
the shape of the sample volume in PW Doppler has a large effect on the content of the doppler signal
true
to measure flow velocity with color flow imaging
a sample volume is placed in the faster streamline and set at angle correction parallel to vessel wall
To form the color flow image, the system samples down each Doppler image scan line for
The average Doppler shift frequency
Separating moving soft tissues from moving blood in a color flow image test for
The relationship between amplitudes and frequencies of the Doppler signal
The smallest vessels that appear in a color flow image are about 1 mm diameter. The smaller vessels are absent because
Small vessels do not reflect as well as larger vessels do
Detecting color flow depends on
Changes in echo signal phase
Power Doppler imaging is a good choice for
Showing blood perfusion through the tissue
Power Doppler is limited because
It does not show direction of flow
Turbulance in color flow vascular imaging appears as
A mottled pattern of color or mottled pattern of red and blue with changing saturation
Color flow imaging in the vascular system usually uses
A linear sequence array transducer
What is the maximum velocity limit for a 3 megahertz continuous wave Doppler image at 4 cm deep
CW has no maximum limit
Aliasing occurs when the frequency shift exceed
1/2 the PRF
When imaging the portal vein in the liver and no color is detected in the vein, what can you do to improve the sensitivity of flow
Decrease PRF (for slow flow)
The packet size ( or ensemble length) in color Doppler refers to the number of
Pulses per scan line
If the Doppler angle of incidence is underestimated, what error will result?
Velocity estimate will be inaccurate
The processing technique for determining direction in CW Doppler is
Fast Fourier Transform
What method is used to steer the color box?
Electronic time delays
If the frequency of the returning echo is higher than the fundamental transmitted frequency, which will occur?
Positive shift with red color
The Doppler spectral display shows
Relative signal power at each frequency in the Doppler signal
If you increase the wall filter during Doppler sampling, it will
Reduce display of low-frequency shift
What technique is used to perform the spectral analysis?
Fast fourier transform
If you increase the transmit frequency of the color Doppler, what will happen?
Improved sensitivity to slow flow
If you lower the color threshold, what will occur?
Increased color fill within the vessel
The Doppler frequency shift is defined as
The difference between transmitted and received frequencies
From the Doppler equation, which would not be true?
Increasing Doppler frequency will incase frequency shift
what does brightness of the z axis on the Doppler spectrum refer to?
Amplitude
The Nyquist limit refers to
Sampling frequency required for Doppler samples without aliasing
What can you adjust to improve the frame rate?
Decrease size of color box and decreased scan line density
What adjustments can you make to decrease aliasing?
Increase PRF
With the following scan parameters, what is the frequency at which aliasing will occur? 5MHz fundamental frequency, 3 kHz PRF, 4 cm depth, 1.5 mm sample volume width.
3 kHz ( Nyquist limit is 1/2 PRF)
The faster techniques used to obtain the mean Doppler frequency in color flow is
Autocorrelation
If you see spectral broadening in the spectral analysis display, it usually means
Turbulent flow
Which of the following refer to the number of pulses used to measure the velocity along one scan line?
Packet size or ensemble length
Which analysis method would show turbulance with color Doppler?
variance mode color display
Which term refers to the axial length of the sample volume
Color gate
Spectral analysis is used for
Showing the distribution and magnitude of frequency shift in the Doppler signal
what adjustments can you make to eliminate color flash artifact?
increase wall filter
Which of the following is not true for turbulent flow?
causes increase in pressure downstream
Which conditions are most likely to produce turbulance?
5 mm diameter with 300 cm / sec velocity (Reynolds number =average speed x diameter x density)
Which of the following are true regarding color Doppler images?
Color image is superimposed over the B mode image
Which of the following will decrease frame rate?
Increasing line density
When using duplex scanning, how many pulses per scan line are usually used to create the gray scale image?
1 - 2
If you increase the packet size, what will result?
Decreased frame rate
Which controls on the control panel adjusts Doppler PRF?
Spectral velocity scale
When you increase the Doppler frequency, what occurs?
Increased Doppler frequency shift
What should you adjust to improve color Doppler sensitivity to flow?
Lower wall filter settings
If you lower the PRF, which of the follow occurs?
Nyquist limit is decreased
What part of the Doppler spectrum in color -encoded with color Doppler
The frequency shift
What color would represent flow at 90 degrees to the sound beam?
Black
What color is usually represented as moving for the transducer?
Red
Why does the frame rate decrease when adding color to the grayscale image?
More pulses are required for each scan line
Which control determines the grayscale brightness at which color will display instead of grayscale?
Priority
With reference to the Doppler equation, the frequency shift will increase by increasing the
Transmit frequency
What effects the spatial resolution of the color image?
Frequency and line density
What happens if you increase the Doppler transmit frequency?
Increased Doppler frequency shift
What happens if you increase the Doppler angle?
Doppler frequency shift decreases
The velocity estimation would have the greatest error at Doppler angle?
70 degrees
Another term for ensemble length or dwell time is
Packet size
Which of the following occur to Doppler frequency shift with high velocity in the vessels?
Increased frequency shift
The appearance of the spectral window on the spectral waveform display is affected by
All of above, turbulent flow, Doppler gain, sample volume size
What must be known to calculate the velocity from the Doppler shift information?
All of the above, Doppler frequency, angle of incidence, speed of sound in tissues
Which control would you use to increase the amplification of the color Doppler signal?
color gain
Which of the following could result in color Doppler display?
Any of the above, blood flow, motion from respiration, cardiac motion, transducer motion
Which control removes low frequency and display a clean zero baseline?
High pass filter
Continuous wave (CW) Doppler
does not alias
Although the signals from CW Doppler are not as complex, only pulsed Doppler can do
Depth location
Which of the following is / are true of pulse Doppler?
Determines depth by echo arrival time
CW Doppler does not alias because
the transducer is continuously sending and receiving with shows the changing flow patterns rather than sampling at interval
Statement is not true of CW Doppler?
Aliasing can occur at the very high frequency
In pulse Doppler, which of the following do not affect the shape and size of the sample volume?
Spectral scale range (this applies only to the processing of the echo that are received)
When measuring flow velocity, the pixel color represents
Movement of both blood and /or tissue
When comparing CW and pulse wave Doppler which apply?
CW has a duty cycle of 1.00
If the returning Echoes have a lower frequency than the transmitted frequency, the reflector is
Moving away from the transducer
If the flow is toward a transducer of 2 MHz, the frequencies of the Doppler shift should be _____ 2MHz
Much less than
If a 5 MHz transducer display a peak shift of 12 kHz, what shift would the same vessel display with a 2.5 MHz transducer
6 kHz
If a Doppler shift of 3 kHz is displayed with an angle of 40 degrees, what will the shift be at 60 degrees?
Less than
If the scanner does not have the ability to display the spectral analysis waveform, it will not show
The distribution of the range of frequencies
The range gate determines
The sample volume length along the length of the beam direction
The frequency of the Doppler samples is determined by the
The pulse repetition
Aliasing is more likely at lower Doppler frequency when the sample is
Farther from the transducer
Adjustments to overcome aliasing do not include
Using a higher fundamental frequency
Range (depth) discrimination is poorest with
CW Doppler
If you increase the packet size (ensemble length), the result is
Better velocity estimation with lower frame rate
The term variance refers to
velocity range (laminar flow has less, turbulent flow has more)
The most accurate measurement of velocity distribution is with
pulsed Doppler
The scanning term that display gray scale and a spectral waveform is called
Duplex scanning