the doppler effect occurs
|
all echoes from moving sources
|
|
the doppler effect depends on
|
the closing velocity between transducer and tissue, carrier frequency, and u/s propagation velocity
|
|
the shape of the sample volume in PW Doppler has a large effect on the content of the doppler signal
|
true
|
|
to measure flow velocity with color flow imaging
|
a sample volume is placed in the faster streamline and set at angle correction parallel to vessel wall
|
|
To form the color flow image, the system samples down each Doppler image scan line for
|
The average Doppler shift frequency
|
|
Separating moving soft tissues from moving blood in a color flow image test for
|
The relationship between amplitudes and frequencies of the Doppler signal
|
|
The smallest vessels that appear in a color flow image are about 1 mm diameter. The smaller vessels are absent because
|
Small vessels do not reflect as well as larger vessels do
|
|
Detecting color flow depends on
|
Changes in echo signal phase
|
|
Power Doppler imaging is a good choice for
|
Showing blood perfusion through the tissue
|
|
Power Doppler is limited because
|
It does not show direction of flow
|
|
Turbulance in color flow vascular imaging appears as
|
A mottled pattern of color or mottled pattern of red and blue with changing saturation
|
|
Color flow imaging in the vascular system usually uses
|
A linear sequence array transducer
|
|
What is the maximum velocity limit for a 3 megahertz continuous wave Doppler image at 4 cm deep
|
CW has no maximum limit
|
|
Aliasing occurs when the frequency shift exceed
|
1/2 the PRF
|
|
When imaging the portal vein in the liver and no color is detected in the vein, what can you do to improve the sensitivity of flow
|
Decrease PRF (for slow flow)
|
|
The packet size ( or ensemble length) in color Doppler refers to the number of
|
Pulses per scan line
|
|
If the Doppler angle of incidence is underestimated, what error will result?
|
Velocity estimate will be inaccurate
|
|
The processing technique for determining direction in CW Doppler is
|
Fast Fourier Transform
|
|
What method is used to steer the color box?
|
Electronic time delays
|
|
If the frequency of the returning echo is higher than the fundamental transmitted frequency, which will occur?
|
Positive shift with red color
|
|
The Doppler spectral display shows
|
Relative signal power at each frequency in the Doppler signal
|
|
If you increase the wall filter during Doppler sampling, it will
|
Reduce display of low-frequency shift
|
|
What technique is used to perform the spectral analysis?
|
Fast fourier transform
|
|
If you increase the transmit frequency of the color Doppler, what will happen?
|
Improved sensitivity to slow flow
|
|
If you lower the color threshold, what will occur?
|
Increased color fill within the vessel
|
|
The Doppler frequency shift is defined as
|
The difference between transmitted and received frequencies
|
|
From the Doppler equation, which would not be true?
|
Increasing Doppler frequency will incase frequency shift
|
|
what does brightness of the z axis on the Doppler spectrum refer to?
|
Amplitude
|
|
The Nyquist limit refers to
|
Sampling frequency required for Doppler samples without aliasing
|
|
What can you adjust to improve the frame rate?
|
Decrease size of color box and decreased scan line density
|
|
What adjustments can you make to decrease aliasing?
|
Increase PRF
|
|
With the following scan parameters, what is the frequency at which aliasing will occur? 5MHz fundamental frequency, 3 kHz PRF, 4 cm depth, 1.5 mm sample volume width.
|
3 kHz ( Nyquist limit is 1/2 PRF)
|
|
The faster techniques used to obtain the mean Doppler frequency in color flow is
|
Autocorrelation
|
|
If you see spectral broadening in the spectral analysis display, it usually means
|
Turbulent flow
|
|
Which of the following refer to the number of pulses used to measure the velocity along one scan line?
|
Packet size or ensemble length
|
|
Which analysis method would show turbulance with color Doppler?
|
variance mode color display
|
|
Which term refers to the axial length of the sample volume
|
Color gate
|
|
Spectral analysis is used for
|
Showing the distribution and magnitude of frequency shift in the Doppler signal
|
|
what adjustments can you make to eliminate color flash artifact?
|
increase wall filter
|
|
Which of the following is not true for turbulent flow?
|
causes increase in pressure downstream
|
|
Which conditions are most likely to produce turbulance?
|
5 mm diameter with 300 cm / sec velocity (Reynolds number =average speed x diameter x density)
|
|
Which of the following are true regarding color Doppler images?
|
Color image is superimposed over the B mode image
|
|
Which of the following will decrease frame rate?
|
Increasing line density
|
|
When using duplex scanning, how many pulses per scan line are usually used to create the gray scale image?
|
1 - 2
|
|
If you increase the packet size, what will result?
|
Decreased frame rate
|
|
Which controls on the control panel adjusts Doppler PRF?
|
Spectral velocity scale
|
|
When you increase the Doppler frequency, what occurs?
|
Increased Doppler frequency shift
|
|
What should you adjust to improve color Doppler sensitivity to flow?
|
Lower wall filter settings
|
|
If you lower the PRF, which of the follow occurs?
|
Nyquist limit is decreased
|
|
What part of the Doppler spectrum in color -encoded with color Doppler
|
The frequency shift
|
|
What color would represent flow at 90 degrees to the sound beam?
|
Black
|
|
What color is usually represented as moving for the transducer?
|
Red
|
|
Why does the frame rate decrease when adding color to the grayscale image?
|
More pulses are required for each scan line
|
|
Which control determines the grayscale brightness at which color will display instead of grayscale?
|
Priority
|
|
With reference to the Doppler equation, the frequency shift will increase by increasing the
|
Transmit frequency
|
|
What effects the spatial resolution of the color image?
|
Frequency and line density
|
|
What happens if you increase the Doppler transmit frequency?
|
Increased Doppler frequency shift
|
|
What happens if you increase the Doppler angle?
|
Doppler frequency shift decreases
|
|
The velocity estimation would have the greatest error at Doppler angle?
|
70 degrees
|
|
Another term for ensemble length or dwell time is
|
Packet size
|
|
Which of the following occur to Doppler frequency shift with high velocity in the vessels?
|
Increased frequency shift
|
|
The appearance of the spectral window on the spectral waveform display is affected by
|
All of above, turbulent flow, Doppler gain, sample volume size
|
|
What must be known to calculate the velocity from the Doppler shift information?
|
All of the above, Doppler frequency, angle of incidence, speed of sound in tissues
|
|
Which control would you use to increase the amplification of the color Doppler signal?
|
color gain
|
|
Which of the following could result in color Doppler display?
|
Any of the above, blood flow, motion from respiration, cardiac motion, transducer motion
|
|
Which control removes low frequency and display a clean zero baseline?
|
High pass filter
|
|
Continuous wave (CW) Doppler
|
does not alias
|
|
Although the signals from CW Doppler are not as complex, only pulsed Doppler can do
|
Depth location
|
|
Which of the following is / are true of pulse Doppler?
|
Determines depth by echo arrival time
|
|
CW Doppler does not alias because
|
the transducer is continuously sending and receiving with shows the changing flow patterns rather than sampling at interval
|
|
Statement is not true of CW Doppler?
|
Aliasing can occur at the very high frequency
|
|
In pulse Doppler, which of the following do not affect the shape and size of the sample volume?
|
Spectral scale range (this applies only to the processing of the echo that are received)
|
|
When measuring flow velocity, the pixel color represents
|
Movement of both blood and /or tissue
|
|
When comparing CW and pulse wave Doppler which apply?
|
CW has a duty cycle of 1.00
|
|
If the returning Echoes have a lower frequency than the transmitted frequency, the reflector is
|
Moving away from the transducer
|
|
If the flow is toward a transducer of 2 MHz, the frequencies of the Doppler shift should be _____ 2MHz
|
Much less than
|
|
If a 5 MHz transducer display a peak shift of 12 kHz, what shift would the same vessel display with a 2.5 MHz transducer
|
6 kHz
|
|
If a Doppler shift of 3 kHz is displayed with an angle of 40 degrees, what will the shift be at 60 degrees?
|
Less than
|
|
If the scanner does not have the ability to display the spectral analysis waveform, it will not show
|
The distribution of the range of frequencies
|
|
The range gate determines
|
The sample volume length along the length of the beam direction
|
|
The frequency of the Doppler samples is determined by the
|
The pulse repetition
|
|
Aliasing is more likely at lower Doppler frequency when the sample is
|
Farther from the transducer
|
|
Adjustments to overcome aliasing do not include
|
Using a higher fundamental frequency
|
|
Range (depth) discrimination is poorest with
|
CW Doppler
|
|
If you increase the packet size (ensemble length), the result is
|
Better velocity estimation with lower frame rate
|
|
The term variance refers to
|
velocity range (laminar flow has less, turbulent flow has more)
|
|
The most accurate measurement of velocity distribution is with
|
pulsed Doppler
|
|
The scanning term that display gray scale and a spectral waveform is called
|
Duplex scanning
|