  • Front
  • Back
Equipment & Instrumentation: Automatic Pressure Injectors
Parts
'Head' where syringes with contrast material are inserted
Plungers push the contrast into tubing attached to catheters
High pressure mechanism - Electromechanical system with motor drive connected to a screw.
Floor or overhead mount
Control unit where flow rate and volumes are programmed
Equipment & Instrumentation: Automatic Pressure Injectors
Function
Administers a precise volume (bolus) of contrast at a rapid preset flow rate
Bolus injections rates for LV-gram, PA, Aorta are 30-60ml at 15-30ml/sec
Watch for bubbles!!
Equipment & Instrumentation: Automatic Pressure Injectors
Operation
Load Syringe
Set Flow Rates
Press trigger to inject/or during cine from table
Safety features:
warns of omitted settings or mechanical malfunction
Volume limiting device
Pressure limiting device: sets max PSI, prevents catheter to whiplash (recoil), breakage, vessel damage
Accelerator regulators: guards against recoil, exact duration of time
Rate rise control: allows for gradual rise in PSI
Equipment & Instrumentation: Automatic Pressure Injectors
Types & Control Panel Info
Air compressed (not used)
Pressure Dependent
Fixed Flow Rate
**
Flow rate - amount per time
Volume - total amount
Linear Rise - time to reach desired flow
PSI - pressure that you inject
Injection delay - inject late
Film changer delay - to inject early