What is the function of any Patient Breathing System (PBS) and what are the types?
|
the function is to deliver O2, +/- anesthetic drug and to elminate CO2
two types -rebreathing -nonrebreathing |
|
What are consequences of leaks in the breathing system?
|
-not enough anesthetic
-not enough O2 -polluting the environment |
|
What is the oxygen flush valve used for?
What does it bypass? What is the effect? |
-Delivers O2 to the Patient Breathing System
-Bypasses the vaporizor -Effects: Dilute the anesthetic |
|
What does the fresh gas inlet supply?
|
gases that have not be re-breathed and where the O2 and the anesthetic originates
|
|
What is the purpose of the pop off ALP valve?
|
Limits the pressure build during manual or spontaneous ventilation
KEEP THIS OPEN!!!! unless needed for manual, controlled or assisted ventilation |
|
What does the breathing system pressure gauge measure?
|
the pressure withing the patient breathing system
this should always read ZERO |
|
What is used to absorb CO2 and in what system?
|
Soda lime in a rebreathing system
|
|
What are the signs in the patient that the CO2 absorber may be exhausted?
|
-increased end tidal CO@ (should be 35-45mm/Hg)
-increased ventilation -rebreathing -respiratory acidosis -increased HR and BP -read mucous membranes |
|
What is tidal volume?
|
the volume of on inspiration and one expiration
|
|
How do you calculate tidal volume?
|
10-20ml/kg body weight
reservoir bag should be ~6x the tidal volume |
|
T/F
The reservoir bah should always be empty. |
false
should be full |
|
What are the uses of the reservoir bag?
|
-watch the respiration of the patient
-Intermittent positive pressure ventilation (IPPV) -inspiratory reservoir |
|
What are the two non rebreathing systems used at Ross University?
|
Mapleson D and F
|
|
What non rebreathing systems at Ross has an APL valve?
|
Mapleson D
|
|
What do we used the non rebreathing system for?
|
patients under 10 kg due to its low resistance and high flow
|
|
What ppm should the exposure to halogenated anesthetics be kept to?
|
2ppm
|
|
What machine collects excess gas from the machine?
|
Scavenging
|
|
What are the passive scavenging systems?
|
-floor drop
-charcoal absorption -window or outside wall -vacuum collection |
|
When is the charcoal absorber replaced?
|
When it weighs more than 50g then it did after 8-12 hours of use
|
|
What are the sources of anesthetic pollution in addition to used gas?
|
-gas sampling monitors
-filling the vaporizors -recovery rooms (exhaling patients) -induction chambers of face masks |
|
What should be turned off before filling a vaporizer with anesthetic?
|
vapor dial
|